176 Chapter 12: System Configuration
TempScan / MultiScan User's Manual
To Compute the MultiScan/1200 Scan Interval
To compute the time interval for the MultiScan/1200 to acquire one scan, use the following formulas:
•
For a 60-Hz Line Cycle:
Scan Interval equals:
Is-m = Cv x (Wv + 12) x R3
•
For a50-Hz Line Cycle:
Scan Interval equals:
Is-m = Cv x (Wv + 12) x R4
Where:
Is-m
= Scan Interval for the MultiScan/1200, milliseconds.
Cv
= Channels, variable number of configured channels in the scan.
Wv
= Sampling Weight, variable number of samples per channel.
R3
= Sampling Rate, constant of 0.52083 milliseconds (per sample) for a 60-Hz line cycle.
R4
= Sampling Rate, constant of 0.62500 milliseconds (per sample) for a 50-Hz line cycle.
For example, if 148 channels are configured, the disposition of these channels will not be a factor.
Using the scan interval formula for a MultiScan/1200 system with a 60-Hz line cycle, if the system is
configured to take 1 sample per channel, gives:
Is-m = 148 x (1 + 12) x 0.52083 millisec.
Is-m = 148 x 13 x 0.52083 millisec = 1002 millisec = 1.002 seconds.
Meanwhile, the following table shows how the
fast mode
scan rate varies when the number of
configured channels increases where these MultiScan/1200 channels are all adjacent to each other, and
are using the default sampling weight of 32 samples per channel.
Variations of the MultiScan/1200 Fast Mode Scan Rate
Number of
Number of
Scan Rate (Hz)
12-Channel
Selected
60-Hertz Line Cycle
50-Hertz Line Cycle
Blocks
Channels
Hertz
seconds
Hertz
seconds
1
12
3.636
0.275
3.030
0.330
2
24
1.818
0.550
1.515
0.660
3
36
1.212
0.825
1.010
0.990
4
48
0.909
1.100
0.758
1.320
5
60
0.727
1.375
0.606
1.650
…
…
…
…
…
…
10
120
0.3636
2.750
0.3030
3.300
…
…
…
…
…
…
50
600
0.0727
13.750
0.0606
16.500
…
…
…
…
…
…
60
720
0.0606
16.500
0.0505
19.800
61
732
0.0596
16.775
0.0497
20.130
62
744
0.0587
17.050
0.0489
20.460
Note:
(1) The number of selected channels refer to configured channels which are all adjacent to
each other. (2) These values were obtained from the above MultiScan/1200 scan interval
equations.
Содержание MultiScan 1200
Страница 2: ...TempScan MultiScan User s Manual...
Страница 4: ...TempScan MultiScan User s Manual...
Страница 10: ...vi TempScan MultiScan User s Manual Notes...
Страница 20: ...10 Chapter 1 System Overview TempScan MultiScan User s Manual Notes...
Страница 106: ...96 Chapter 8 ChartView ChartView Plus TempScan MultiScan User s Manual Notes...
Страница 130: ...120 Chapter 9 TempView TempScan MultiScan User s Manual...
Страница 150: ...140 Chapter 9 TempView TempScan MultiScan User s Manual Notes...
Страница 166: ...156 Chapter 10 PostView TempScan MultiScan User s Manual Notes...
Страница 234: ...224 Chapter 13 System Operation TempScan MultiScan User s Manual Notes...
Страница 248: ...238 Chapter 14 System Calibration TempScan MultiScan User s Manual Notes...
Страница 284: ...274 Chapter 15 Program Examples TempScan MultiScan User s Manual Notes...
Страница 330: ...320 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Страница 342: ...332 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Страница 358: ...348 Appendix TempScan MultiScan User s Manual Notes...
Страница 364: ...354 Index TempScan MultiScan User s Manual Notes...