328 Chapter 16: API Command Reference
TempScan / MultiScan User's Manual
T
- Set Trigger Configuration
TYPE
Acquisition
EXECUTION
Deferred
SYNTAX
Tstart,
stop,
re-arm
,
sync
Configure an acquisition of scan data with the following parameters:
start
is the start event and
stop
is the stop event
re-arm
determines if the unit should continue acquiring after the first
acquisition completes, where
1
is on and
0
is off.
sync
determines if acquisition events should be synchronized with the
internal timebases, where
1
is on and
0
is off.
T?
Query the present trigger configuration
DESCRIPTION
The
Set Trigger Configuration
(
T
) command is used as the central command to assign the following events to the
TempScan/1100 or MultiScan/1200 unit: The
Trigger
(trigger start event) and the
Stop
(trigger stop event). The
Trigger
and
Stop
may also be assigned different trigger sources. In addition, the
Set Trigger Configuration
(
T
)
command also determines whether or not, after the initial acquisition, the acquisition will be re-enabled
automatically, and whether or not the
Trigger
(trigger start event) should be synchronized with the
Pre-Trigger
scan interval.
Since the
Set Trigger Configuration
(
T
) command governs the configuration of an acquisition, all other
acquisition-dependent commands should be sent before the trigger sources are assigned via the
Set Trigger
Configuration
(
T
) command. The required acquisition-dependent commands will depend on the selected trigger
sources. For instance, if the unit is configured for
Trigger On
(
@
) command, it will not be necessary to send the
unrelated
Set Trigger Level
(
L
) command.
The following table lists the valid trigger types and their associated commands prior to trigger configuration:
Required Commands Prior To
Trigger Configuration
Trigger
Type
Trigger Source Description
With Pre-
Trigger/ Post-
Stop Scans
Without Pre-
Trigger/ Post-
Stop Scans
Optional
Commands
Prior To Trigger
Configuration
0
None - Event Not Defined
(N/A)
(N/A)
(N/A)
1
Trigger on (
@
) Command
I,Y
I
(N/A)
2
Trigger on GET (IEEE 488 only)
I,Y
I
(N/A)
3
Trigger on TALK (IEEE 488 only)
I,Y
I
(N/A)
4
Trigger on Channel Above Level
I,Y,L
I,L
(N/A)
5
Trigger on Channel Below Level
I,Y,L
I,L
(N/A)
6
Trigger on TTL Signal Rising
I,Y
I
(N/A)
7
Trigger on TTL Signal Falling
I,Y
I
(N/A)
8
Trigger on Count
I,Y
I,Y
(N/A)
9
Trigger on Alarm Turning On
I,Y,C
I,C
A
10
Trigger on Alarm Turning Off
I,Y,C
I,C
A
11
Trigger on Absolute Time
I,Y,P
I,P
(N/A)
The first parameter
start
defines the
Trigger
(trigger start event) that is to take place in order for the unit to
begin acquiring
Post-Trigger
scans. Setting the
start
parameter to
0
will disable the acquiring of
Post-Trigger
scans; this may be performed at any time during or after an acquisition. The
start
parameter will automatically
be set to
0
upon the termination of an acquisition unless the
re-arm
flag is set to
1
.
The second parameter
stop
defines the
Stop
(trigger stop event) that is to take place in order for the unit to stop
acquiring
Post-Trigger
scans. Setting the
stop
parameter to
0
will have the same effect as setting the
stop
parameter to type
8
(Trigger on Count) with the
Post-Stop
count (
Y
command) set to
0
. In both cases, the
Post-
Trigger
scans will be collected and the acquisition immediately terminated.
The third parameter
re-arm
defines the
Auto Re-arm
feature which determines whether successive
acquisitions will be re-enabled automatically, or remain disabled after the initial acquisition has terminated.
The fourth parameter
sync
defines the trigger synchronization flag which determines whether the
Trigger
will be
synchronized with the
Pre-Trigger
scan interval (
I
command), or begin as soon as the
Trigger
is detected,
regardless of the exact time it occurred.
Содержание MultiScan 1200
Страница 2: ...TempScan MultiScan User s Manual...
Страница 4: ...TempScan MultiScan User s Manual...
Страница 10: ...vi TempScan MultiScan User s Manual Notes...
Страница 20: ...10 Chapter 1 System Overview TempScan MultiScan User s Manual Notes...
Страница 106: ...96 Chapter 8 ChartView ChartView Plus TempScan MultiScan User s Manual Notes...
Страница 130: ...120 Chapter 9 TempView TempScan MultiScan User s Manual...
Страница 150: ...140 Chapter 9 TempView TempScan MultiScan User s Manual Notes...
Страница 166: ...156 Chapter 10 PostView TempScan MultiScan User s Manual Notes...
Страница 234: ...224 Chapter 13 System Operation TempScan MultiScan User s Manual Notes...
Страница 248: ...238 Chapter 14 System Calibration TempScan MultiScan User s Manual Notes...
Страница 284: ...274 Chapter 15 Program Examples TempScan MultiScan User s Manual Notes...
Страница 330: ...320 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Страница 342: ...332 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Страница 358: ...348 Appendix TempScan MultiScan User s Manual Notes...
Страница 364: ...354 Index TempScan MultiScan User s Manual Notes...