TempScan / MultiScan User's Manual
Chapter 15: Program Examples 243
MultiScan/1200
Example 15d. HLL48.BAS Program
(1)
CLS : PRINT “The MultiScan/1200 is resetting...”
PRINT #1, “OUTPUT 07;*RX”
(3)
SLEEP 4
(4)
WHILE (S% AND 4) = 0
PRINT #1, “SPOLL 07"
INPUT #2, S%
(7)
WEND
(8)
PRINT #1, “OUTPUT 07;M#0X”
(9)
PRINT #1, “OUTPUT 07;W#32X”
(10)
PRINT #1, “OUTPUT 07;C1-24,3X”
(11)
PRINT #1, “OUTPUT 07;C25-48,14X”
(12)
PRINT #1, “OUTPUT07;Y1,0,0X”
(13)
PRINT #1, “OUTPUT07;T1,1,0,0X"
(14)
PRINT “The MultiScan/1200 is collecting HLL data for all 48 channels”
PRINT “Hit a key to start or stop retrieving HLL data...”
(16)
WHILE INKEY$ = “”: WEND
(17)
WHILE INKEY$ = “”
PRINT #1, “OUTPUT 07;U4X”
PRINT #1, “ENTER 07"
LINE INPUT #2, U$
(21)
FOR i = 1 TO 24
PRINT “High, Low and Last readings of Channel”; i
PRINT MID$(U$, (i * 66) - 65, 66)
NEXT i
(25)
FOR i = 1 TO 24
PRINT “High, Low and Last readings of Channel”; i + 24
PRINT MID$(U$, (24 * 66) + (i * 78) - 77, 78)
NEXT i
(29)
WEND
This program is
HLL48.BAS
in the
\EXAMPLES\
subdirectory. It will configure 24 thermocouple and
24 volts channels, collect the High/Low/Last (HLL) Registers data, and post it on the screen.
•
Lines 1-3:
Although not necessary, it is good practice to reset the MultiScan/1200 at the start of
your application by sending the
Reset
Power-On
(
*R
) command. Wait a few seconds.
•
Lines 4-7:
To be certain that the reset was successful and the device is ready, Serial Poll the device
until the proper status is returned.
•
Lines 8-9:
The first thing necessary to start scanning is to setup the measuring mode. In this case,
the MultiScan/1200 will be setup for line-cycle integration / high-speed multi-channel mode.
•
Lines 10-11:
The
Configure Channels
(
C
) command will be used once to configure a range of
channels from 1 to 24 as Type T thermocouples, and used again to configure a range of channels
from 25 to 48 as
±
10 volts DC inputs.
•
Lines 12-13:
Program for one
Pre-Trigger
scan via the
Set Counts
(
Y
) command. Then program
the
Trigger
(trigger start event) and
Stop
(trigger stop event) via the
Set Trigger Configuration
(
T
)
command to arm the acquisition. Telling the MultiScan/1200 to start acquiring data is the last
thing necessary to start scanning. Since the MultiScan/1200 does not scan channels until an
acquisition is armed, arming an acquisition with one
Pre-Trigger
scan allows valid HLL data to be
read.
•
Lines 14-16:
The next three lines tell you to hit a key to start the acquisition. The program will
not proceed until a key is pressed.
Содержание MultiScan 1200
Страница 2: ...TempScan MultiScan User s Manual...
Страница 4: ...TempScan MultiScan User s Manual...
Страница 10: ...vi TempScan MultiScan User s Manual Notes...
Страница 20: ...10 Chapter 1 System Overview TempScan MultiScan User s Manual Notes...
Страница 106: ...96 Chapter 8 ChartView ChartView Plus TempScan MultiScan User s Manual Notes...
Страница 130: ...120 Chapter 9 TempView TempScan MultiScan User s Manual...
Страница 150: ...140 Chapter 9 TempView TempScan MultiScan User s Manual Notes...
Страница 166: ...156 Chapter 10 PostView TempScan MultiScan User s Manual Notes...
Страница 234: ...224 Chapter 13 System Operation TempScan MultiScan User s Manual Notes...
Страница 248: ...238 Chapter 14 System Calibration TempScan MultiScan User s Manual Notes...
Страница 284: ...274 Chapter 15 Program Examples TempScan MultiScan User s Manual Notes...
Страница 330: ...320 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Страница 342: ...332 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Страница 358: ...348 Appendix TempScan MultiScan User s Manual Notes...
Страница 364: ...354 Index TempScan MultiScan User s Manual Notes...