54 Chapter 7: Power & Assembly
TempScan / MultiScan User's Manual
Power-Up Activation
At initial power-up or on the
Reset Power-On
(
*R
) command, the TempScan/1100 or MultiScan/1200
performs automatic self-tests to ensure that it is fully functional. The front panel LED indicators show
any errors if they occur. Possible error conditions and their corresponding LED indicator patterns are
shown in the following table. Any LED pattern not shown is an internal error that is not field-
serviceable; in this case, contact the factory. If ERROR is on by itself, there is a configuration error
due to setup information in NV-RAM. Check the error using the
Query Error Status
(
E?
) command.
LED Indicators
Error Condition
TALK
LISTEN
SRQ
ERROR
POWER
No Errors
(Off)
(Off)
(Off)
(Off)
ON
General Hardware Failure
ON
(Off)
ON
FLASHING
ON
Position-U22 ROM Invalid
(Off)
(Off)
(Off)
FLASHING
ON
Position-U21 ROM Checksum Error
(Off)
(Off)
ON
FLASHING
ON
Position-U22 ROM Checksum Error
(Off)
ON
(Off)
FLASHING
ON
Non-Volatile RAM Error
ON
(Off)
(Off)
FLASHING
ON
Dynamic RAM Error
ON
ON
(Off)
FLASHING
ON
Interprocessor COM Error
ON
ON
ON
FLASHING
ON
If no problems are found, the
POWER
LED indicator will remain on while the rest of the indicators will
go out, and the TempScan/1100 or MultiScan/1200 will begin its power-up initialization. This self test
is performed each time the unit is powered up regardless of whether power-on was caused by the power
switch or the
Reset Power-On
(
*R
) command. During initialization, the self test performs the following
steps:
1.
Checks for errors at power-up.
2.
Checks the flag in the NV-RAM to determine if it should power-up with factory default settings or
a user-defined configuration.
3.
Loads appropriate registers with corresponding values in NV-RAM.
4.
Checks a flag to see if alarms should be enabled at power-up, and if so, enables them.
5.
Loads channel configuration registers.
6.
Loads program sequencer with appropriate channel configurations.
7.
Resets computations processor to begin acquiring scans.
The self test takes approximately five seconds to complete, after which the TempScan/1100 or
MultiScan/1200 unit is ready for normal operation.
For more information, see section
Power-Up Configuration
in the chapter
System Configuration
.
Содержание MultiScan 1200
Страница 2: ...TempScan MultiScan User s Manual...
Страница 4: ...TempScan MultiScan User s Manual...
Страница 10: ...vi TempScan MultiScan User s Manual Notes...
Страница 20: ...10 Chapter 1 System Overview TempScan MultiScan User s Manual Notes...
Страница 106: ...96 Chapter 8 ChartView ChartView Plus TempScan MultiScan User s Manual Notes...
Страница 130: ...120 Chapter 9 TempView TempScan MultiScan User s Manual...
Страница 150: ...140 Chapter 9 TempView TempScan MultiScan User s Manual Notes...
Страница 166: ...156 Chapter 10 PostView TempScan MultiScan User s Manual Notes...
Страница 234: ...224 Chapter 13 System Operation TempScan MultiScan User s Manual Notes...
Страница 248: ...238 Chapter 14 System Calibration TempScan MultiScan User s Manual Notes...
Страница 284: ...274 Chapter 15 Program Examples TempScan MultiScan User s Manual Notes...
Страница 330: ...320 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Страница 342: ...332 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Страница 358: ...348 Appendix TempScan MultiScan User s Manual Notes...
Страница 364: ...354 Index TempScan MultiScan User s Manual Notes...