Channels
Number of Slots:
One (1) slot
Number of Channels (TempScan/1100):
Up to 32 differential thermocouple or voltage inputs, or up
to 16 RTD inputs; accepts TempTC/32B, TempV/32B, or TempRTD/16B scanning modules
Number of Channels (MultiScan/1200):
Up to 24 differential thermocouple or voltage inputs;
accepts MTC/24 or MHV/24 scanning modules.
Channel Attributes:
High and low set points; hysteresis values for high and low set points.
Scan Sequence:
Any combination of temperature and voltage channels may be scanned, but channels
are scanned in ascending numerical order.
Scan Interval:
Absolute time between scans (
hh:mm:ss.s
); min = 00:00:00.0, max = 99:59:59.9.
Note that specifying a value of 00:00:00.0 results in no delay between channel scans.
Scanning Modes (TempScan/1100):
For thermocouples up to 500 feet:
960 channels/sec @ 60 Hz;
800 channels/sec @ 50 Hz;
For thermocouples over 500 feet:
240 channels/sec @ 60 Hz, 200
channels/sec @ 50 Hz.
Scanning Modes (MultiScan/1200):
For multi-channel scanning:
147 channels/sec @ 50 or 60 Hz;
For 32-point line-cycle averaging enabled:
44 channels/sec @ 60 Hz, 38.5 channels/sec @ 50 Hz;
For single-channel burst mode:
1 channel @ 20K samples/sec.
Triggers
Installation Category:
For CE:
Category 1.
Programmable Triggering:
Temperature or Voltage level (above or below), absolute time of day,
alarm condition (on or off), IEEE GET, IEEE TALK, external TTL trigger (rising or falling),
specified number of readings.
Temperature-Level Trigger:
Programmable value for any one channel.
For MultiScan/1200:
This
trigger not available when in single-channel burst mode.
TTL Trigger:
Programmable for rising or falling edges.
Pre-Trigger Count:
Programmable integer (< memory size -1).
Post-Trigger Count:
Programmable integer.
Trigger Input Connector:
External BNC connector
Trigger Output Connector:
External BNC connector
Содержание MultiScan 1200
Страница 2: ...TempScan MultiScan User s Manual...
Страница 4: ...TempScan MultiScan User s Manual...
Страница 10: ...vi TempScan MultiScan User s Manual Notes...
Страница 20: ...10 Chapter 1 System Overview TempScan MultiScan User s Manual Notes...
Страница 106: ...96 Chapter 8 ChartView ChartView Plus TempScan MultiScan User s Manual Notes...
Страница 130: ...120 Chapter 9 TempView TempScan MultiScan User s Manual...
Страница 150: ...140 Chapter 9 TempView TempScan MultiScan User s Manual Notes...
Страница 166: ...156 Chapter 10 PostView TempScan MultiScan User s Manual Notes...
Страница 234: ...224 Chapter 13 System Operation TempScan MultiScan User s Manual Notes...
Страница 248: ...238 Chapter 14 System Calibration TempScan MultiScan User s Manual Notes...
Страница 284: ...274 Chapter 15 Program Examples TempScan MultiScan User s Manual Notes...
Страница 330: ...320 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Страница 342: ...332 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Страница 358: ...348 Appendix TempScan MultiScan User s Manual Notes...
Страница 364: ...354 Index TempScan MultiScan User s Manual Notes...