Operation - Calibration
Fusion4 MSC-L
Part No.: 4418309_Rev09
5 - 276
Installation & Operation Manual
Honeywell
9. On the
Calibration . Stream n . Wizard 5/6
screen, select
<OK>
on
the IR controller or LAD, to accept the new meter factor and view the
calibration details.
The
Calibration . Stream n . Wizard 6/6
screen appears, which
displays the calibration details.
A new calibration record is created and stored in the system. Each
calibration is saved in a non-volatile memory, with date, time, old and
new values.
NOTE: Select
<ESC>
to restore the old meter factor.