Operation - Calibration
Fusion4 MSC-L
Part No.: 4418309_Rev09
5 - 254
Installation & Operation Manual
Honeywell
FIGURE 5-35
Calibrating the flow meter for an Additive Stream
With the flow meter and the K-factor value, a correction factor can be
calculated, which is used for (re-)calibrating the flow meter.
This correction factor is called the meter factor.
The resulting injection volume (V) is then calculated as follows:
V = Number of pulses/(K-factor * meter factor).