Operation - Calibration
Part No.: 4418309_Rev09
Fusion4 MSC-L
Honeywell
Installation & Operation Manual
5 - 255
5.18.2 Calibration Menu Choice
1. On the
Main Menu
screen, select the
Calibration
icon to perform
the calibration process.
FIGURE 5-36
Calibration icon
2. On the
Calibration
screen, select either
<Wizard>
calibration method
or the
<Manual>
calibration method.
In the
<Wizard>
calibration method, the meter factor is calculated
by comparing the measured volume with the device dispensed
volume.
In the
<Manual>
calibration method, you must enter the meter
factor.
The default meter factor for the
<Wizard>
calibration method and the
<Manual>
calibration method is 1. The range which can be specified
for the meter factor is between 0.5 and 1.5.