109
C90
Saving the adjustments data in the PO’s
Options
:
0
:
No storage
1
:
Data is saved in the permanent memory.
Special function
:
None.
Description:
The calibration data determined with C80, C81, C83, C84, C88, C115, C122 can be transferred to the
permanent memory using code C90.
To this end use the option selection and set “1”. Then acknowledge with the Acknowledgment key.
The data which used to be temporary to this moment is now transferred and saved in the permanent memory.
Saving is acknowledged by a series of audible signals.
Comments:
None
C92
Display of actual distance and diameter of inner SAPE
Options
:
None
Special function
:
None
Description:
1. Field / left display
: diameter (408-410)
2. Field / right display
: distance (From machine to calibration rotor edge).
Comments:
None.
C97
Conditioning of the solenoid brake
Options
:
None
Special function
:
None
Description:
Conditioning of the solenoid brake (p-variants only)
WARNING!
With this test function the main shaft still rotates when the wheel guard is open.
Enter code C97 and then press the START key to run the motor with brake activated in order to remove contamination
such as fi ngerprints from the brake lining, thus improving the braking effect of a new solenoid brake. As this function
causes the brake to heat up quickly, it is deactivated automatically after one minute.
The motor speed is held constant by controlling the brake current.
Comments:
Brake current is not read out.
C98
Display of angular position of power clamp pulley, incremental encoder test
Options :
Special function:
Resetting incremental register fl ags.
Description:
Once this code is called up, the reading is “---”
Only after the incremental encoder has identifi ed zero reference, the angular location is read out in a range between 0
and 63.
Use of the C code and meaning of the fl ags are identical with C code C74 (incremental encoder of main shaft).
Comments:
If this test fails (no 23F) please check
•
the cabling of the Optoencoder – micro-controller and
•
the cabling Optoencoder power clamp - Optoencoder main shaft
•
the connectors of the cables
•
clean the incremental encoder tape
SM HOF GEODYNA 7600 Rel.D