64
0X001021
IEM ZERO MISMATCH
Check wing nut
If the problem still persists:
Check the shaft.
If the problem still persists:
Check the wheel clamping.
If the problem still persists:
Check the main shaft Optoencoder fl at cable.
If the problem still persists:
Check the main shaft Optoencoder.
If the problem still persists:
Check the IBP box
If the problem still persists:
Call Snap On
C Codes
C74, C75, C85, C86, C115, C84, C88, C90
0X001022
IEP ZERO MISMATC
Check the derive pulley Optoencoder adjustment.
If the problem still persists:
Check the derive pulley Optoencoder.
If the problem still persists:
Check if the drive pulley moves side to side.
If the problem still persists:
Call Snap On
C Codes
C98
0X001023
SPEED FLUCTUATION DURING THE MEASUREMENT
Check the belt tension.
If the problem still persists:
Check the speed with C60.
If the problem still persists:
Check Optoencoder with C54.
If the problem still persists:
Check the capacitor
If the problem still persists:
Call Snap On
C Codes
C28, C54, C60
0X001024
NOISY SIGNAL FROM TRANSDUCERS DURING MEASURING RUN
Check if the machine is well placed on the fl oor.
If the problem still persists:
Check if the balancer components vibrate.
If the problem still persists:
Check the belt tension.
If the problem still persists:
Check the bearings.
If the problem still persists:
Check the transducers
If the problem still persists:
Check if the vibratory assu is well secured to the cabinet.
If the problem still persists:
Call Snap On
C Codes
C28
SM HOF GEODYNA 7600 Rel.D