Install potentiometer onto bracket and tighten 13mm nut.
Ø
Install cog wheel onto potentiometer shaft and hand. Tighten nut.
Ø
Attach SAPE harness to Processor Board.
Ø
Turn on the machine
Ø
Run service code C80, C81, C82, C90 for SAPE calibration.
Ø
Mount the weight tray.
Ø
Check if the machine works fine.
4.25 LASER POINTER: CHECK AND REPLACEMENT
"
:
Medium and big philip screwdriver, 5mm allen key.
6
: 1h
i
: Defective laser pointer may cause the following malfunctions:
1. Turning the machine on several error 50X are displayed.
2. The laser does not work at all.
3. Laser shoots the dot in one point only.
TO CHECK THE LASER POINTER
Ø
Turn on the balancer.
Ø
Check with C123.
Ø
Turn of the balancer and disconnect it from power supply.
Ø
Remove the weight tray.
Ø
Check if the cable and connector are fi ne.
TO REPLACE THE LASER POINTER
Ø
Turn off the balancer.
Ø
Remove the weight tray of the balancer.
Ø
Disconnect the laser cables from the weight laser board.
Ø
Remove the laser plastic cover.
Ø
Remove the two screws that secure the laser assembly to the VPM.