111
C110
Indication of the operating voltages supplied by the power supply module
Options
:
None.
Special function
:
None.
Description:
Indication of the operating voltage, 5.0 VDC, supplied by the power supply module for the controller
board
Comments:
Please refer to error ID 810 and 811 for the limits.
C111
Belt tension: Measure fi rst harmonic of the belt
Options:
FOR STEEL VPM ONLY
Special function:
None.
Description:
Once this code is called up, the reading is a running “ - “.
After beating on the belt a beep indicates the start of the measurement. After 3 sec. a second beep indicates the end
of the measurement and the frequency is displayed. If the harmonic analysis doesn’t fi nd any signifi cant frequency,
“Err” is displayed until the next measurement is started.
The test should be repeated with different forces during picking the belt and with different sections of the belt (rotating
the main shaft).
The frequency is displayed in a range from 100 Hz up to 300 Hz.
The begin of a new measurement will be detected automatically.
Comments:
Only for belt driven balancer.
C115
Calibration of the unbalance measurement with test rotor.
Options
:
None.
Special function
:
None.
Description:
With the calibration of the unbalance measurement the following are determined:
-
the sensitivity of the transducers,
-
the phase difference of the transducer signals,
-
the comparative data for readjustment by the operator and temperature compensation
-
the phase shift of the unbalance signal amplifi ers and
-
the angular deviation
After the 1
st
step, that is the measuring run, a beep signal is heard.
After acknowledgement/setting of weight size in step 2 a beep signal is heard (in addition to the beep made by the
key).
In step 6 the ambient transducer temperature will be read out for 1 second.
NOTE: THIS TEST REQUIRES THE USE OF PRUEFROTOR
. ALSO CHECK THE VCC VOLTAGE “C110” IF
NECESSARY BEFORE ANY CALIBRATION IS DONE.
Mount the calibration tool on the balancer shaft . Set the calibration tool data.
SM HOF GEODYNA 7600 Rel.D