101
Step 4:
Move gauge head to calibration position 4. Press acknowledge key to confi rm and quit the calibration.
Position 2 of calibration template, or tip
(screw) of calibration weight
Comments:
Save calibration data in the permanent memory using C90 code
C81
Measuring the adaptor fl ange and the zero plane
Options
:
None
Special function
:
None
Description:
To keep the machine operative and to allow width measurement in case the
gauge arm is defective, the distance SAPE will be calibrated relative to the
clamping surface of the adaptor on the condition that the gauge arm has already
been calibrated. Having applied the gauge arm to the outer edge of the adaptor
fl ange, press the Acknowledgment key to acknowledge the input. No further
inputs are necessary.
Comments:
Save the calibration data in the the permanent memory using
code C90.
C82
Calibration of the outer SONAR
Options :
None
Special function:
None
Description:
Before performing the C82, it is mandatory to perform C80 and C81.
1. Mount the calibration rotor with the bracket downward at 6 o’clock.
2. Position the wings of the nut horizontally in order to avoid any
interference with the sonar.
SM HOF GEODYNA 7600 Rel.D