107
Step 1:
Remove wheel adaptor from main shaft.
Clamp the medium and big cones on the shaft and lock
Close the wheel guard and press "START" key for the
measurement run.
Step 2:
Unclamp the clamping sleeve and place the calibration ring between
the cone and the sleeve. Lock the cones again.
Step 3:
Close the wheel guard and press "START" key for the measurement run.
Comments:
Complete this calibration with C88 and C90
C85
Copy content of serial EEPROM (EEP) from IBP to optoencoder
Options
:
0
:
No action
1
:
Copy from IBP box EEP to opto EEP.
Special function
:
None
Description:
To make sure the content is not overwritten by mistake, fi rst set display from 0 to 1 by using the
option selection
.
Then the copying operation is started by pressing the Acknowledgment key.
Once copying has been completed, the machine is started again automatically.
Comments:
None
C86
Copy content of serial EEPROM (EEP) from optoencoder to IBP
Options
:
0
:
No action
1
:
Copy from optoencoder EEP to IBP box EEP.
Special function
:
None.
Description:
To make sure the content is not overwritten by mistake, fi rst set display from 0 to 1 by using the
option selection.
Then the copying operation is started by pressing the Acknowledgment key.
Once copying has been completed, the machine is started again automatically.
Comments:
See C85
SM HOF GEODYNA 7600 Rel.D