0X001057
CAL1 0G WEIGHT HIGH
Perform the calibration with C115.
If the problem still persists:
Check the transducers wiring
If the problem still persists:
Check the transducers adjustment.
If the problem still persists:
Check the IBP box.
If the problem still persists:
Call Snap On
C Codes
C66, C75, C103, C104, C115, C84, C88, C90.
0X001058
CAL2 0G WEIGHT HIGH
Perform the calibration with C115.
If the problem still persists:
Check the transducers wiring
If the problem still persists:
Check the transducers adjustment.
If the problem still persists:
Check the IBP box.
If the problem still persists:
Call Snap On
C Codes
C66, C75, C103, C104, C115, C84, C88, C90.
0X001059
C1 USER CAL TOOL LOW
Perform the User calibration.
If the problem still persists:
Check the User Calibration weight.
If the problem still persists:
Check the transducers wiring
If the problem still persists:
Check the transducers.
If the problem still persists:
Check the IBP box.
If the problem still persists:
Call Snap On
C Codes
C66, C75, C103, C104, C115, C84, C88, C90
0X001060
C2 USER CAL TOOL LOW
Perform the User calibration.
If the problem still persists:
Check the User Calibration weight.
If the problem still persists:
Check the transducers wiring
If the problem still persists:
Check the transducers.
If the problem still persists:
Check the IBP box.
If the problem still persists:
Call Snap On
C Codes
C66, C75, C103, C104, C115, C84, C88, C90