3.2 Configuring Measurement Settings
54
About the Recording Interval
• Select the data acquisition interval to suit your measurement objectives.
• Note that shorter recording intervals restrict the maximum recording time.
• The minimum recording interval depends on the input unit (channels),
and whether burn-out detection is enabled.
With burn-out detection (Burn Out) disabled (Off)
When channels 1 to 15 of UNIT1 are selected for measurement: 10 ms
When channels 1 to 15 of UNIT2 are selected for measurement: 20 ms
When channels 1 to 15 of UNIT3 and UNIT4 are selected for measure-
ment: 50 ms
With burn-out detection (Burn Out) enabled (On)
When channels 1 to 15 of UNIT1 are selected for measurement: 20 ms
When channels 1 to 15 of UNIT2 are selected for measurement: 50 ms
When channels 1 to 15 of UNIT3 and UNIT4 are selected for measure-
ment: 100 ms
See:
"Chapter 11 Specifications" (p. 227)
• Some waveform peaks may not be displayable with certain interval set-
tings.
For example, when the
[Interval]
is 1 s and the
[Time/DIV]
(time base) is
5 s/DIV.
See:
About Horizontal Axis Settings: (p. 70)
• With the 50 or 60 Hz digital filter setting, longer recording intervals pro-
duce lower cut-off frequencies and better noise suppression (p. 75).
• The recording interval cannot be changed while measuring. Stop mea-
surement before setting.
During Normal Recording (
[Cont]
disabled (Off), and
[Repeat]
enabled
(On))
After a recording length has elapsed, some “dead time” is needed for inter-
nal processing before recording can resume. No measurement occurs dur-
ing the dead time. So to avoid information loss, split data into files at a
specified interval by enabling continuous recording (
[Repeat]
), and select-
ing
[Split Save]
for auto saving.
5s/div
1s
Peak
Dead time
Measuring
Содержание LR8400-20
Страница 1: ......
Страница 2: ......
Страница 4: ......
Страница 26: ...Operating Precautions 14...
Страница 96: ...3 9 Viewing and Editing with the All Channel Settings List 84...
Страница 112: ...4 3 Marking Waveforms and Searching Marks Search Function 100...
Страница 156: ...6 5 Transferring Data to a PC USB Drive Mode 144...
Страница 238: ...10 8 About Communications Commands 226...
Страница 256: ...11 3 LR8501 Universal Unit Specifications 244...
Страница 262: ...12 3 Disposing of the HiLogger 250...
Страница 296: ...Index Index 4...
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......