3.3 Input Channel Settings
64
Configure these settings for each pulse channel receiving input from a pulse output
device such as a watt-hour or flow meter.
Make these settings on the
[CH]
screen.
See:
"Key Setting Procedure" (p .55)
Integration (Count) Measurement Settings
1
Select the input unit (
PLS&ALM
) and channel
(
P1
to
P8
), and check the box to enable the
channel.
2
Select
[Count]
.
3
Select the integration mode (Count Mo).
4
Select the basis of measurement (Slope).
5
Select the High/Low reference (threshold) val-
ues.
6
Select a filter setting.
Setting options:(
∗
: default setting)
Add
∗
Counts the number of pulses acquired from
the start of measurement.
Inst
Counts the number of pulses acquired dur-
ing each recording interval. The pulse count
is reset at the beginning of each recording
interval.
Setting options:(
∗
: default setting)
↑∗
Counts the number of L - H transitions.
↓
Counts the number of H - L transitions.
Setting options:(
∗
: default setting)
1V
∗
Judge 1.0 V or higher as HIGH, and 0 to
0.5 V as LOW.
4V
Judge 4.0 V or higher as HIGH, and 0 to
1.5 V as LOW.
Setting options:(
∗
: default setting)
Off
∗
Filter is disabled.
On
Filter is enabled. Prevents incorrect counts
due to relay contact chatter (noise).
Totalized
[Add]
Instantaneous
[Inst]
0 V
L
H
1
: On
: Off
2
3
4
5
6
• Use the scaling function to display integrated pulse counts as the cor-
responding physical units (such as Wh or VA) of the measurement
parameter (p. 71).
• The counter saturates at 1,073,741,823 counts. If you expect to
exceed this count, we recommend measuring with the Integration
Mode (Count Mo) set to Instantaneous (Inst), and summing later with
a spreadsheet program.
Содержание LR8400-20
Страница 1: ......
Страница 2: ......
Страница 4: ......
Страница 26: ...Operating Precautions 14...
Страница 96: ...3 9 Viewing and Editing with the All Channel Settings List 84...
Страница 112: ...4 3 Marking Waveforms and Searching Marks Search Function 100...
Страница 156: ...6 5 Transferring Data to a PC USB Drive Mode 144...
Страница 238: ...10 8 About Communications Commands 226...
Страница 256: ...11 3 LR8501 Universal Unit Specifications 244...
Страница 262: ...12 3 Disposing of the HiLogger 250...
Страница 296: ...Index Index 4...
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......