6.2 Saving Data
128
1
Select the contents to be saved.
If
[Off]
is selected, skip steps 2 to 5.
If
[Calc (post meas.)]
is selected, skip steps 4 and 5.
2
Specify the waveform file name (if needed).
See:
"3.7 Entering Titles and Comments (as needed)"
(p. 73)
If left blank, file names are created automatically.
See:
"Appendix 3 File Naming" (p. A8)
3
Select the removable storage priority (CF card or
USB flash drive)
4
Select whether to delete while saving.
Select the desired operation when removable storage be-
comes full while saving.
5
Select whether to save data into multiple files
(“Split Save”).
Select single (per measurement) or multiple (timed) file
creation.
6
Press
START
to start measuring.
After measurement, the data is automatically saved to re-
movable storage (p. 123).
Setting options: (
∗
: default setting)
Off
∗
, Waveform(realtime), CSV(realtime), Calc (post
meas.), Wa Calc, CSV + Calc
Setting options: (
∗
: default setting)
Off
∗
Saving ends when the removable storage is
nearly full.
On
When the free capacity of the removable stor-
age reaches a certain minimum size, wave-
form files are deleted, oldest first. If deletion is
not possible, saving stops. Delete while saving
does not apply to numerical calculation files.
Setting options: (
∗
: default setting)
Off
∗
Creates a single file per measurement.
On
Specify the save timing (split time). (Figure at
right) A new data file is created for each split
time (span).
Ref
Time
Specify the reference time (Ref Time) and split
time (Split Length). (Figure at right) A new file
is created for each split time (span), starting
from the reference time.
1
2
3
4
With
[Split Save]
set to
[On]
With
[Split Save]
set to
[Ref Time]
Also set the following items:
[Ref Time]
: 0:0 to 23:59
[Split Length]
:1, 2, 5, 10, 15,
20, 30 minutes, 1, 2, 3, 4, 6, 8,
12 hours, 1 day
Also set the
[Split Length]
.
Days (0 to 30), Hours (0 to 23),
Minutes (0 to 59)
Before measurement starts,
"Sync fixed time" is displayed
until the next measurement
time synchronizes with the ref-
erence time.
The horizontal axis setting is lim-
ited when auto saving waveforms
(p. 70).
5
Before starting to measure,
confirm that Auto Save is
configured correctly, and
that the removable storage
is properly installed.
Содержание LR8400-20
Страница 1: ......
Страница 2: ......
Страница 4: ......
Страница 26: ...Operating Precautions 14...
Страница 96: ...3 9 Viewing and Editing with the All Channel Settings List 84...
Страница 112: ...4 3 Marking Waveforms and Searching Marks Search Function 100...
Страница 156: ...6 5 Transferring Data to a PC USB Drive Mode 144...
Страница 238: ...10 8 About Communications Commands 226...
Страница 256: ...11 3 LR8501 Universal Unit Specifications 244...
Страница 262: ...12 3 Disposing of the HiLogger 250...
Страница 296: ...Index Index 4...
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......