3.9 Viewing and Editing with the All-Channel Settings List
81
3
3
Chapt
e
r
3 Sett
ings
A value measured on CH1 of UNIT1 (Analog channel CH1-1) can be set as the scaling
value. This function is convenient for thermocouple measurements when errors (such as
thermocouple deviations) give different values for the same known temperature.
It applies only to channels with the same input type and range setting as CH1 of UNIT1.
Setting CH1 of UNIT1 Value as a Scaling Value (Inter-Channel
Compensation function)
When scaling is automatically set by steps 1 to 6 on next page procedure,
the measurement units are initially set to [V].
To set all channels to
[
°
C]
make the following settings beforehand.
1. Set the
[Convert]
setting to
[Dec]
and
[unit]
to
[
°
C]
on the
[Conv/Calc]
screen.
2. For
[Batch Proc]
, execute
[Correct the value to the same as CH1
when input type is same.]