3.6 Scaling Settings (as needed)
71
3
3
Chapt
e
r
3 Sett
ings
Use the scaling function to convert input volt-
age to the physical units of the measurement
parameter for display, such to convert voltage
input for display as electrical current.
Converted values can be displayed in fixed or
floating-point notation.
Make these settings on the
[CH]
screen.
See:
"Key Setting Procedure" (p .67)
3.6
Scaling Settings (as needed)
Normal Display
(Off)
Scaling Enabled
A
mV
The Setting screen cannot be dis-
played while measuring.
Scaling Voltage, Temperature, Humidity, Resistance, and Revolution
Measurements
1
Select the desired numerical notation.
2
Select the scaling method.
:
3
Enter the
[Unit]
name on the comment entry
screen. (Up to 7 characters)
See:
About Comments (p. 73)
4
Ratio: Enter the conversion ratio and offset.
2-pt: Enter the values at two points before and
after conversion.
Move the cursor to the numerical value and press
EN-
TER
to display the numerical setting dialog.
See:
"Numerical Value Entry" (p .69)
Setting options:(
∗
: default setting)
Dec
Displays converted values using fixed-point
notation.
Exp
Displays converted values using floating-
point notation.
Off
∗
Scaling is disabled.
Setting options:(
∗
: default setting)
Ratio
∗
Convert by specifying the ratio of physical
units per volt of input signal (conversion ra-
tio), offset, and the name of the units.
2-pt
Convert by specifying the input signal volt-
age at two points, the physical unit values at
those points, and the name of the units.
1 2
3
Conversion Ratio Setting
2-Point Setting
2-Point Setting
With
[Cond]
set to
[Exp]
With
[Cond]
set to
[Dec]
4
Conversion Ratio Setting
A value measured on CH1 of UNIT1 (Analog channel CH1-1) can be set as the scaling
value (Inter-Channel Compensation function) (p. 81).
Содержание LR8400-20
Страница 1: ......
Страница 2: ......
Страница 4: ......
Страница 26: ...Operating Precautions 14...
Страница 96: ...3 9 Viewing and Editing with the All Channel Settings List 84...
Страница 112: ...4 3 Marking Waveforms and Searching Marks Search Function 100...
Страница 156: ...6 5 Transferring Data to a PC USB Drive Mode 144...
Страница 238: ...10 8 About Communications Commands 226...
Страница 256: ...11 3 LR8501 Universal Unit Specifications 244...
Страница 262: ...12 3 Disposing of the HiLogger 250...
Страница 296: ...Index Index 4...
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......