Appendix 2 Error Messages and Remedial Actions
A
4
239
CSV auto save has been set.Un-
able to set the specified record in-
terval.
Select non-CSV auto saving to measure with the
specified recording interval.
240
The timer stop time has been
passed.
Change the timer stop time or cancel the timer.
241
10ms recording interval cannot be
set due to too many channels.
Measurement with 10 ms recording interval is avail-
able when using UNIT1-1 to -15. Disable (set to Off)
channels on other input units (2 to 4).
242
20ms recording interval cannot be
set due to too many channels.
Measurement with 20 ms recording interval is avail-
able when using UNIT1-1 to UNIT2-15. Disable (set
to Off) channels on other input units (3 and 4).
243
Recording interval has been
changed.
The recording interval has changed due to expand-
ed channel usage.
244
Auto-save has been changed from
CSV to waveform.
The saving method was changed to Waveform (bi-
nary) because the selected recording interval is too
short for CSV saving. To enable CSV saving, select
a longer recording interval or reduce the number of
channels used.
See:
"Automatic Saving (Waveform Data and
Numerical Calculation Results)" (p .127)
∗
2
301
No waveform data.
No waveform data is preset, so data saving and cal-
culations cannot be performed. Press
START
to
start acquiring measurement data.
302
Confirm the A-B cursor position.
The A/B cursor positions are invalid (out of wave-
form range). Check the cursor positions.
303
No numerical calculation data.
Execute numerical calculation
304
Invalid event mark.
Try moving to an event mark that is not in memory.
401
File processing error.
An unexpected fault occurred while processing a file
on the removal storage. Turn the HiLogger off and
back on.
402
Cannot load this file.
The format of the file is incompatible with the HiLog-
ger, or the file is corrupt.
Refer to
403
Insert a Media.
A CF card or USB flash drive is not present. Insert
storage media.
404
Directory full.
No more files can be created because the directory
is full. Check or perform the following:
1. System reset (p. 153)
2. Verify that the removal storage is specified by
3. Reformat the removal storage (p. 49)
If the message persists after performing the above,
either the removal storage or HiLogger may be dam-
aged. Contact your dealer or Hioki representative.
405
Not enough capacity.
The removal storage as insufficient space to save
the file. Delete files to increase free space, or re-
place the removal storage.
Message
Remedial Action
Содержание LR8400-20
Страница 1: ......
Страница 2: ......
Страница 4: ......
Страница 26: ...Operating Precautions 14...
Страница 96: ...3 9 Viewing and Editing with the All Channel Settings List 84...
Страница 112: ...4 3 Marking Waveforms and Searching Marks Search Function 100...
Страница 156: ...6 5 Transferring Data to a PC USB Drive Mode 144...
Страница 238: ...10 8 About Communications Commands 226...
Страница 256: ...11 3 LR8501 Universal Unit Specifications 244...
Страница 262: ...12 3 Disposing of the HiLogger 250...
Страница 296: ...Index Index 4...
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......