5.2 Alarm Output
115
3
5
Chapter 5 S
p
ecif
y
ing Crite
ria
for M
easurement
s
An alarm can be output when thermocouple burn-out is detected.
Setting options: (
∗
: default setting)
OR
∗
Alarm output occurs when the alarm criteria
specified for any alarm-enabled channel are
satisfied.
AND
Alarm output occurs only when the alarm crite-
ria for every alarm-enabled channel are satis-
fied.
3. Selecting Combined Alarm Output Criteria
Select the
[Trig & Alm]
Screen.
If
[Alarm]
is set to
[Off]
,
change it to (
[On]
).
1
Select the
[Trig & Alm]
Screen.
2
Select a
[Pulse Logic Alarm]
unit.
3
Set
[Alarm at burn out. ]
to
[On]
.
This setting causes alarm output to be generated whenever a broken wire is detected,
regardless of other alarm condition settings and alarm source settings
[AND, OR]
.
1
2
3
Содержание LR8400-20
Страница 1: ......
Страница 2: ......
Страница 4: ......
Страница 26: ...Operating Precautions 14...
Страница 96: ...3 9 Viewing and Editing with the All Channel Settings List 84...
Страница 112: ...4 3 Marking Waveforms and Searching Marks Search Function 100...
Страница 156: ...6 5 Transferring Data to a PC USB Drive Mode 144...
Страница 238: ...10 8 About Communications Commands 226...
Страница 256: ...11 3 LR8501 Universal Unit Specifications 244...
Страница 262: ...12 3 Disposing of the HiLogger 250...
Страница 296: ...Index Index 4...
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......