Appendix 1 Scan Timing
A
1
Appendix
L
For sampling of the LR8500 Voltage/Temp Unit and LR8501 Universal Unit, there is
a scan system for each channel.
In the following figures, the digital filter is Off, recording interval is 20 ms (with 2
input units), or 50 ms (with 4 input units).
When the digital filter is off, the scan widths and dead times between channels are
fixed to the values shown in the figure above regardless of the recording intervals.
When the digital filter is on, the scan widths and dead times between channels vary
in accordance with the recording intervals.
Appendix
Appendix 1 Scan Timing
With Two Input Units (max. 20 ms sampling)
Unit 1
Unit 2
10 ms
10 ms
10 ms
Data Refresh
With Four Input Units (max. 50 ms sampling)
Unit 1
Unit 2
10 ms
10 ms
10 ms
Unit 3
Unit 4
10 ms
10 ms
10 ms
Data Refresh
No input channels are connected dur-
ing this period.