background image

 

 

SPS Smart Profiler 

Hardware Guide 

 
 
 
 
 
 

 

 
 
 
 
 
 
 

Publication Number PTG-330310-000 Rev. C 

 
 
 
 

 

 

Содержание Smart Dock

Страница 1: ...SPS Smart Profiler Hardware Guide Publication Number PTG 330310 000 Rev C...

Страница 2: ...reserved Patents are pending This document and the information contained in it may not be sold duplicated used or disclosed in whole or in part except as specifically authorized in writing by KIC The...

Страница 3: ...ED BEHAVIOR 6 THE POWER BUTTON 6 SET UP THE SPS SMART PROFILER 7 CHOOSING THE MODE OF PROFILING 8 DATALOG MODE 8 WIRELESS MODE 11 HARDWARE STATUS LED STATUS WINDOW 14 USING THE THERMAL SHIELD 15 SAFE...

Страница 4: ...n and faster cooldown between profiles The design of the SPS thermal shields allows for easy and secure opening and closing durability that meets the most stringent of drop tests and temperature toler...

Страница 5: ...r cutting the tape Gloves for handling the profiler and test product hot from the thermal process SPS Smart Profiler Getting Started Guide SPS Smart Profiler CD containing Profiling Software 2G SPS Sm...

Страница 6: ...e software The normal setup routine must follow the sequence shown below 1 Install the software 2 Install the USB security dongle 3 Connect the SPS Smart Profiler hardware to the computer Note When co...

Страница 7: ...2 3 Alert Not Ready Solid On Cannot profile due to TC 1 missing TCs too hot SPS too hot Battery too low startup failure voltage too low to charge or memory full Plug into PC and software to identify a...

Страница 8: ...ng Data Collection Profiling Fast Flashing Profiling in progress Profile Complete Flashing every 3 seconds One or more profiles are complete and stored in SPS memory Data Empty Off No profiles stored...

Страница 9: ...le Battery LED Behavior LED Name Icon Color Mode LED Behavior Mode Description Battery Charging Combined solid and flashing See description Plugged into USB or AC charger for charging Each LED represe...

Страница 10: ...es through the thermal process that it later will transfer to the computer through a cable Note Once the Profiling Software 2G is installed and running whenever you connect an SPS Smart Profiler to a...

Страница 11: ...Your Profiling Software USB Security Dongle is required to use all features in the software 2 Use the communication cable to connect the SPS Smart Profiler to a USB port on the computer 3 A message a...

Страница 12: ...isplays radio button choices for specifying the communication mode 6 Click the radio button for Datalogger followed by the Save green check button 7 Click OK on the message boxes no messages appear if...

Страница 13: ...the status of the profiler and confirming that the software is connected to the SPS Smart Profiler and receiving live readings Note The Profiler panel displays live temperature values and the current...

Страница 14: ...Use the communication cable to connect the SPS Smart Profiler to a USB port on the computer 3 A message appears confirming the powered connection 4 From the Main screen click on the Globe button to d...

Страница 15: ...be returned to the main screen 8 Remove the cable from the SPS and insert it into the Smart Dock 9 Tap the SPS on the Smart Dock to pair it 10 The ANT blue LED on both devices flashes until the devic...

Страница 16: ...g the status of the profiler and confirming that the software is receiving the live Wi Fi signal from the SPS Smart Profiler Note The Profiler panel displays live temperature values and the current ba...

Страница 17: ...attery indicates percent of battery charge Alert if LED is on will list any and all alerts in the list box Data LED status will indicate the number of profiles currently stored in the SPS with access...

Страница 18: ...The SPS Smart Profiler s maximum internal temperature is 85 C ALWAYS use the thermal shield to protect the SPS Smart Profiler from heat damage Note Gloves are required when the profiler has just emer...

Страница 19: ...tion Never substitute gloves that are not rated for similar temperature and holding cycle specifications Contact your profiler manufacturer for replacement or extra pairs Safely Remove the Profiler fr...

Страница 20: ...ear the safety gloves supplied with your profiler kit during this procedure 1 Carefully pull up on the latch to open the shield 2 Carefully lift the profiler out of the shield 3 If desired you can pla...

Страница 21: ...ervoltage Category 1 Caution The SPS Smart Profiler measuring circuits are intended for use with non energized equipment test samples only Do not plug in and attach thermocouples to anything that is e...

Страница 22: ...trie Canada exempts de licence standard RSS s Son fonctionnement est soumis aux deux conditions suivantes 1 cet appareil ne doit pas provoquer d interf rences et 2 cet appareil doit accepter toute int...

Страница 23: ...process Note The maximum temperature setting is the highest temperature that the SPS is set to read The highest temperature that the profiler is capable of reading is 1050 C 1922 F This profiler is f...

Страница 24: ...s displayed in the Device field the proper number of channels is identified the appropriate temperature scale is selected and the proper filtering for your location is selected Note When the software...

Страница 25: ...ight influence your selection of this value You can type in whatever value better matches your specific situation If your thermal process is well established around specific temperatures it may make s...

Страница 26: ...software may indicate other values as the default temperature 3 Adjust the output on the thermocouple simulator to match your specified calibration temperature 4 When the values match click the Start...

Страница 27: ...om KIC Service Repair All service and repair of the SPS Smart Profiler or Smart Dock other than the calibration requires contacting a KIC Service center and having the service repair conducted by pers...

Отзывы: