SPS Smart Profiler Hardware Guide
PTG-330310-000 Rev. C
18
Specifications and normal environmental conditions
Accuracy:
±0.5°C
Resolution:
±0.1°C
Internal Operating Temp:
0°C to 85°C
Humidity Range:
20-85% non-condensing
Measurement Range:
-150°C to 1050°C
Sample Rate:
Up to 50 readings per second
Data Points:
72000 per channel
PC Connection:
USB 2.0 (Std-A/Micro-B)
Power:
NiMH battery: 2.4Vdc
USB connection to PC:
5Vdc, 500mA
Wireless Communication:
Wi-Fi2.4 GHz and ANT 2.4 GHz
Thermocouple Compatibility:
7, 9, and 12-Channel Models
Type K, Standard
Dimensions (L x W x H mm):
7-Channel Model:
188.0 x 60.0 x 17.0
9-Channel Model:
188.0 x 75.0 x 17.0
12-Channel Model:
188.0 x 98.0 x 17.0
For Indoor Use Only
Overvoltage Category 1
Caution: The SPS Smart Profiler measuring circuits are intended for use with non-energized
equipment/test-samples only. Do not plug in and attach thermocouples to anything that is
energized.
Max constant voltage handled: 3.15v
Max transient voltage handled: 60v for less than 1ms
Cleaning Procedure
Clean the exterior surfaces of the chassis with a dry lint-free cloth. If any dirt remains, use a cloth or swab dipped in
a 75% isopropyl alcohol solution. Use a swab to clean narrow spaces around the button and connector. Do not use
abrasive compounds on any part of the chassis, as they may damage it.
Avoid the use of chemical cleaning agents, which might damage the plastics used in this instrument. Use a 75%
isopropyl alcohol solution as a cleaner and wipe with a clean cloth dampened with deionized water. Before using
any other type of cleaner, consult your KIC representative.