2.4 Making Connections
37
3
2
Chapter 2 M
eas
ureme
n
t Prep
a
rat
ions
<Connecting to the terminals of the Universal Unit>
Required items: phillips screwdriver (for M3 screws), flat-blade screwdriver (tip width 2.6
mm), measurement leads (for voltage measurement), or thermocouple
(for temperature measurement)
Recommended wire diameter:Single strand diameter:
φ
0.4 mm -
φ
1.2 mm (AWG26-16)
Multi-strand:
0.2 mm
2
- 0.75 mm
2
(AWG24-20)
Standard insulation stripping length: 8 mm
1
With a Phillips screwdriver,
loosen the screws in the termi-
nal block cover, and lift the
cover.
2
Push in the button on the con-
nector with a flatblade screw-
driver.
3
With the button held in, insert
the cable into the cable con-
nection hole.
6
Replace the terminal block
cover, and tighten the screws.
Torque screws to no more than
0.6 N
•
m.
2
3
4
4
Release the button.
The cable is locked.
button
5
Connect to the measurement
object.
5
The insulation colors of wires, which are to be
connected to the plus and minus terminals, differ
depending on countries or manufactures. Check
with the manufacturer on the colors and connect
them according to instructions.
Содержание LR8400-20
Страница 1: ......
Страница 2: ......
Страница 4: ......
Страница 26: ...Operating Precautions 14...
Страница 96: ...3 9 Viewing and Editing with the All Channel Settings List 84...
Страница 112: ...4 3 Marking Waveforms and Searching Marks Search Function 100...
Страница 156: ...6 5 Transferring Data to a PC USB Drive Mode 144...
Страница 238: ...10 8 About Communications Commands 226...
Страница 256: ...11 3 LR8501 Universal Unit Specifications 244...
Страница 262: ...12 3 Disposing of the HiLogger 250...
Страница 296: ...Index Index 4...
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......