6.2 Saving Data
131
3
Chapter 6 Savi
ng & Loadi
n
g Dat
a
6
Settings are displayed. Select contents to be saved.
Media
Select the priority save destination when
[Quick Save]
is selected.
Save
Type
Select the type of data to save.
Calculation must be configured before saving numerical calculation results (p. 155).
Format
(When
[Waveform]
is selected) Select the data format.
Data saved in CSV format cannot be reloaded into the HiLogger or Logger Utility pro-
gram. Fortunately, data saved in binary format can be later converted to text (CSV) for-
mat with the HiLogger or Logger Utility.
Span
(When
[Waveform]
is selected) Select the time span to be saved.
• Before saving a partial waveform, specify the span to be saved (p. 92).
• When a save span is specified for calculation, the results are saved for the
specified span.
Executing a Save
Press
SAVE
to immediately save data according to current settings.
When you select [Quick Save] (to save data upon pressing SAVE)
Setting options: (
∗
: default setting)
CF
∗
Save to the CF card first
USB Memory
Save to the USB flash drive first
Setting options: (
∗
: default setting)
Waveform
∗
Save waveform data.
Screen Image
Save screen image.
Setup File
Save setting configuration data.
Calc Results
Save numerical calculation results.
Setting options: (
∗
: default setting)
Binary
∗
Select this format to be able to reload the saved data into the
HiLogger, or so that it can be loaded into the Logger Utility
program.
CSV
Select this format to save the data as a text file. This format
is ideal for loading into spreadsheet programs like Excel.
Setting options: (
∗
: default setting)
All
∗
Save all measured waveform data.
A-B
Save waveform data between A/B cursors.
Start-A
Save waveform data from the start of measurement to cursor A.
Start-B
Save waveform data from the start of measurement to cursor B.
A-End
Save waveform data from cursor A to the end of measurement.
B-End
Save waveform data from cursor B to the end of measurement.
Содержание LR8400-20
Страница 1: ......
Страница 2: ......
Страница 4: ......
Страница 26: ...Operating Precautions 14...
Страница 96: ...3 9 Viewing and Editing with the All Channel Settings List 84...
Страница 112: ...4 3 Marking Waveforms and Searching Marks Search Function 100...
Страница 156: ...6 5 Transferring Data to a PC USB Drive Mode 144...
Страница 238: ...10 8 About Communications Commands 226...
Страница 256: ...11 3 LR8501 Universal Unit Specifications 244...
Страница 262: ...12 3 Disposing of the HiLogger 250...
Страница 296: ...Index Index 4...
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......