Setting Example
Connecting one HiLogger to one PC
When connecting multiple HiLoggers to a PC through a hub
This example is a local area network with no external connection. We recommend using
only private IP addresses. As below, make sure that no host names or IP addresses are
duplicated.
First HiLogger
Second HiLogger
Third HiLogger
Common Settings
Host Name
LOGGER
DHCP
Off
IP Address
192.168.1.2
Subnet Mask
255.255.255.0
Port
880X
Gateway
Off
DNS
Off
Host Name
LOGGER1
IP Address
192.168.1.2
Host Name
LOGGER2
IP Address
192.168.1.3
Host Name
LOGGER3
IP Address
192.168.1.4
DHCP
Off
Subnet Mask
255.255.255.0
Gateway
Off
Port
880X
LAN communications with a PC require that the HiLogger is first set cor-
rectly.
Содержание LR8400-20
Страница 1: ......
Страница 2: ......
Страница 4: ......
Страница 26: ...Operating Precautions 14...
Страница 96: ...3 9 Viewing and Editing with the All Channel Settings List 84...
Страница 112: ...4 3 Marking Waveforms and Searching Marks Search Function 100...
Страница 156: ...6 5 Transferring Data to a PC USB Drive Mode 144...
Страница 238: ...10 8 About Communications Commands 226...
Страница 256: ...11 3 LR8501 Universal Unit Specifications 244...
Страница 262: ...12 3 Disposing of the HiLogger 250...
Страница 296: ...Index Index 4...
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......