Following are examples of typical trigger settings.
5.6
Trigger Setting Examples
Intended Measurement Objective
Ref. No.
(next table)
Acquire data from when you press
START
until you press
STOP
No.1
Acquire data for one minute after each time you press
START
No.2
Acquire data at one-minute intervals for sixty minutes after you press
START
No.3
When the temperature measured on CH1 exceeds 500
°
C, acquire data until you press
STOP
.
No.4
Acquire data from when you press
START
until the temperature measured on CH1 exceeds
500
°
C.
No.5
Acquire data from when the temperature measured on CH1 exceeds 500
°
C until it drops
below 300
°
C.
No.6
Acquire data whenever the temperature measured on CH1 exceeds 500
°
C, until it drops
below 300
°
C.
No.7
Acquire data for only one minute from when the temperature measured on CH1 exceeds
500
°
C.
No.8
Acquire data for one minute before and after the time when the temperature measured on
CH1 exceeds 500
°
C.
No.9
Acquire data from 9:00 to 17:00 on 6/17/2008.
No.10
Starting at 9:00 on 6/17/2008, acquire data for 24 hours/day, for one month.
No.11
Starting on 6/17/2008, acquire data continuously from 9:00 to 17:00 daily for one month.
No.12
Starting on 6/17/2008, acquire data for one hour at 9:00, 15:00, 21:00, and 3:00 daily for one
month.
No.13
Содержание LR8400-20
Страница 1: ......
Страница 2: ......
Страница 4: ......
Страница 26: ...Operating Precautions 14...
Страница 96: ...3 9 Viewing and Editing with the All Channel Settings List 84...
Страница 112: ...4 3 Marking Waveforms and Searching Marks Search Function 100...
Страница 156: ...6 5 Transferring Data to a PC USB Drive Mode 144...
Страница 238: ...10 8 About Communications Commands 226...
Страница 256: ...11 3 LR8501 Universal Unit Specifications 244...
Страница 262: ...12 3 Disposing of the HiLogger 250...
Страница 296: ...Index Index 4...
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......