5.1 Triggering Measurement Start and Stop
108
To use an external signal as a trigger source, make the following settings.
Using External Triggering
1
Connect the external trigger signal to the
HiLogger's EXT.TRIG external control termi-
nal. (See the connection procedure on P.45 )
2
Press
SET
to open the
[System]
screen.
3
Set the
[External Trig In]
to
[Trigger]
(default
setting).
4
Press
SET
to open the
[Trig & Alm]
screen.
5
Set triggering criteria (p. 107).
6
Select the slope for the
[Ext trig] [Start]
and
[Stop]
settings.
Select whether triggering occurs on the trigger sig-
nal's upslope or downslope.
This setting is forced Off when
[External Trig In]
is
set to
[Event]
on the
[System]
screen.
7
Confirm whether triggering is provided by an
external signal or by shorting the EXT.TRIG
terminal to GND (only when
[Ext trig]
is set to
[
↓
]
).
Setting options: (
∗
: default setting)
Off
∗
External triggering is disabled.
↑
Triggering occurs on the upslope from Low
(0 to 1.0 V) to High (2.5 to 5.0 V).
↓
Triggering occurs on the downslope from
High (2.5 to 5.0 V) to Low (0 to 1.0 V), or
when the EXT TRIG terminal is connected
to a GND terminal.
6
3
2
4
6
Содержание LR8400-20
Страница 1: ......
Страница 2: ......
Страница 4: ......
Страница 26: ...Operating Precautions 14...
Страница 96: ...3 9 Viewing and Editing with the All Channel Settings List 84...
Страница 112: ...4 3 Marking Waveforms and Searching Marks Search Function 100...
Страница 156: ...6 5 Transferring Data to a PC USB Drive Mode 144...
Страница 238: ...10 8 About Communications Commands 226...
Страница 256: ...11 3 LR8501 Universal Unit Specifications 244...
Страница 262: ...12 3 Disposing of the HiLogger 250...
Страница 296: ...Index Index 4...
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......