3.3 Input Channel Settings
62
Make these channel settings for resistance measurement.
Make these settings on the
[CH]
screen.
See:
"Key Setting Procedure" (p .55)
Resistance Measurement Settings
1
Select the input unit (
UNIT1
to
4
) and channel
(
1-1
to
4-15
), and check the box to enable the
channel.
2
Select
[Resist]
.
3
Select the appropriate measurement ampli-
tude range.
Setting options:(
∗
: default setting)
10
Ω
∗
, 20
Ω
, 100
Ω
, 200
Ω
1
2
3
: On
: Off
For inductive loads such as coil windings, the slow response time may pre-
vent correct resistance measurement. In this case, select a recording time
longer than 5 s.
When a measured value far exceeds the range limit (range label)
"OVER" or "-OVER" is displayed in place of numerical/cursor values, and the following values
are applied to calculations and data storage.
Input type
Input range
+OVER value -OVER value
Resistance
10
Ω
16.3835
-16.384
20
Ω
32.767
-32.768
100
Ω
163.835
-163.84
200
Ω
327.67
-327.68
Содержание LR8400-20
Страница 1: ......
Страница 2: ......
Страница 4: ......
Страница 26: ...Operating Precautions 14...
Страница 96: ...3 9 Viewing and Editing with the All Channel Settings List 84...
Страница 112: ...4 3 Marking Waveforms and Searching Marks Search Function 100...
Страница 156: ...6 5 Transferring Data to a PC USB Drive Mode 144...
Страница 238: ...10 8 About Communications Commands 226...
Страница 256: ...11 3 LR8501 Universal Unit Specifications 244...
Страница 262: ...12 3 Disposing of the HiLogger 250...
Страница 296: ...Index Index 4...
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......