11.1 LR8400-20, LR8401-20, LR8402-20 Memory HiLogger Specifications
232
(2) Measurement Functions
∗
1: Settable for up to 15 channels (U1-1 to -15) when thermocouple burn-out is disabled.
∗
2: Settable for up to 30 channels (up to U2-1 to U2-15) when thermocouple burn-out disabled, or up
to 15 channels (U1-1 to U1-15) when enabled.
∗
3: Settable for up to 60 channels (up to U3-1 to -15 and U4-1 to -15) when thermocouple burn-out
is disabled, or up to 30 channels (U2-1 to -15) when enabled.
Measurement
Recording interval
10ms
∗
1
, 20ms
∗
2
, 50ms
∗
3
, 100ms, 200ms, 500ms, 1s, 2s, 5s, 10s, 20s,
30s, 1min, 2min, 5min, 10min, 20min, 30min, 1h
Recording length (time
span)
Set continuous recording Off or On.
When Off, any time can be set.
Repeating Measure-
ment Recording
Set Off or On.
When On, measurement repeats at the set recording interval.
Recording Data Quantity
Analog recording n channels: 8 M/n data values
Waveform storage
The most recent 8 million data values (one analog channel recording
time, or 8 M / n values for n channels) are stored in internal memory (n =
number of analog ch pulse channels × 2 + digital channels × 1 +
alarm channels × 5 words)
Data retained in memory can be observed by backward scrolling.
Measurement data
backup
Data is retained for about 30 minutes after power-off
(except when powered off while writing to removable storage, or within
five minutes after power-on)
Display
Display Screen Selec-
tion
(Hybrid display config-
urations)
• Simultaneous waveform, gauge and setting display
• Waveform display
• Simultaneous waveform and numerical value display
• Simultaneous numerical value and comment display
• Numerical value display
• Simultaneous waveforms and numerical calculation results display
• Waveform and A/B cursor values displayed at the same time
Waveform colors
24 colors
Waveform compres-
sion and magnification
Horizontal axis
(time axis)
100 ms, 200 ms, 500 ms, 1 s, 2 s, 5 s, 10 s, 20 s, 30 s, 1 min, 2 min,
5 min, 10 min, 20 min, 30 min, 1 h, 2 h, 5 h, 10 h, 12 h, 1 day /divisions
Vertical axis
(voltage axis)
Select by position or upper/lower limits
• When selecting by position, magnification can be × 100, × 50, × 20, ×
10, × 5, × 2, × 1, or × 1/2. 0 Position: Set between -50% and 150%
(at 1X magnification)
• When selected by upper/lower limits, specify the limit values.
Waveform scrolling
Time-axis scrolling is available by left/right arrow keys while measuring
and when measurement stops (waveform drawing period).
These keys make large jumps
Jump function
Selects the displayed span of the waveform.
Monitor function
Confirm instantaneous values and waveforms without recording data.
(Monitoring is performed while triggering.)
Содержание LR8400-20
Страница 1: ......
Страница 2: ......
Страница 4: ......
Страница 26: ...Operating Precautions 14...
Страница 96: ...3 9 Viewing and Editing with the All Channel Settings List 84...
Страница 112: ...4 3 Marking Waveforms and Searching Marks Search Function 100...
Страница 156: ...6 5 Transferring Data to a PC USB Drive Mode 144...
Страница 238: ...10 8 About Communications Commands 226...
Страница 256: ...11 3 LR8501 Universal Unit Specifications 244...
Страница 262: ...12 3 Disposing of the HiLogger 250...
Страница 296: ...Index Index 4...
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......