11.1 LR8400-20, LR8401-20, LR8402-20 Memory HiLogger Specifications
231
3
Cha
p
te
r 1
1
S
p
ec
ifica
tions
11
Alarm Output
Alarm Output Channels
Four, non-isolated (Common GND with HiLogger)
Output settings
Output occurs when the logical sum (OR) or product (AND) of alarm trig-
ger sources is true.
Alarm source
Analog input
30 channels (U1-1 to U2-15), or up to 60 channels (U1-1 to U4-15) with
additional input units
Integrated pulse in-
put
8 channels (P1-P8)
Digital input
8 channels (L1-L8)
Thermocouple burn-
out
When thermocouple burn-out detection is enabled
Alarm types
analog* and pulse
Level
An alarm is output by input rising above or falling below a preset level
Window
An alarm is output by input crossing into or out of a window defined by
preset upper and lower trigger levels
Trigger types digital
Pattern Mask Alarm
Alarm when signals match the pattern mask (1, 0, and X)
(where X is "ignore")
Alarm level resolution
(Analog) 0.1% f.s. (f.s. = 10 divisions)
(Pulse) Integration 1c, revolution 1/n (where n = pulses/rev)
Latch setting
Alarm output can be set to be held (latched), or not.
Cancel Hold
Enable Alarm Hold to latch alarm output without stopping measurement.
Beeper
Built-in beeper On/Off
Output form
Open-collector output (with 5 V pull-up, active low)
Maximum switch rating
5 to 30 VDC, 200 mA
Output refresh
At every recording interval
Содержание LR8400-20
Страница 1: ......
Страница 2: ......
Страница 4: ......
Страница 26: ...Operating Precautions 14...
Страница 96: ...3 9 Viewing and Editing with the All Channel Settings List 84...
Страница 112: ...4 3 Marking Waveforms and Searching Marks Search Function 100...
Страница 156: ...6 5 Transferring Data to a PC USB Drive Mode 144...
Страница 238: ...10 8 About Communications Commands 226...
Страница 256: ...11 3 LR8501 Universal Unit Specifications 244...
Страница 262: ...12 3 Disposing of the HiLogger 250...
Страница 296: ...Index Index 4...
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......