Appendix 2 Error Messages and Remedial Actions
A
3
Appendix
Appears just once when an error occurs. Disappears within a few seconds. Also
disappears when any key is pressed.
Warning Messages
Message
Remedial Action
101
Invalid key.
The pressed key is disabled due to measuring or
other operation. Press the key again after the cur-
rent operation finishes.
102
Cannot start measurement.
The
START
key is disabled when the current screen
is displayed. Switch to the Waveform/Numerical
screen before starting measurement.
103
Cannot change while measuring.
Press
STOP
to stop measuring before changing the
setting.
201
Exceeding the setting range.
The entered numerical value is out of the valid
range. Enter a valid value.
204
Measurement range changed.
When
[Type]
-
[Up/Lwr]
is selected, the measure-
ment range is automatically changed to match the
upper and lower limits.
230
Recording time has been changed
due to memory limit.
Increasing the number of measurement channels
has decreased Internal memory waveform
data capacity.
231
Pre-Trigger span changed.
Increasing the recording interval or length requires
reducing pre-trigger recording time.
232
Recording and Pre-Trigger spans
changed due to waveform memo-
ry limit.
Increasing the recording interval requires reducing
recording length and pre-trigger recording time.
233
Trigger or alarm changed.
The setting ranges for count integration and revolu-
tions have been changed.
234
Burn Out (broken-wir) detection
changed.
Burn-out detection is not available with the current
settings. To enable burn-out detection, select a long-
er recording interval or reduce the number of chan-
nels used.
See:
"About the Recording Interval" (p .54)
235
Cannot change Burn Out (broken-
wire) detection.
The setting is not available due to the combination of
the number of channels used and the recording in-
terval. To enable burn-out detection, select a longer
recording interval, or reduce the number of channels
used.
See:
"About the Recording Interval" (p .54)
236
File partition period was changed.
The segment time span has been changed in re-
sponse to changing the recording interval.
237
File partition time was changed.
The segment time span has been changed in re-
sponse to changing the recording interval.
238
CSV auto save has been set.Un-
able to set the specified record in-
terval.
The recording intervals available for CSV auto sav-
ing are limited. Select a longer recording interval.
See:
"Automatic Saving (Waveform Data and
Numerical Calculation Results)" (p .127)
∗
2
Содержание LR8400-20
Страница 1: ......
Страница 2: ......
Страница 4: ......
Страница 26: ...Operating Precautions 14...
Страница 96: ...3 9 Viewing and Editing with the All Channel Settings List 84...
Страница 112: ...4 3 Marking Waveforms and Searching Marks Search Function 100...
Страница 156: ...6 5 Transferring Data to a PC USB Drive Mode 144...
Страница 238: ...10 8 About Communications Commands 226...
Страница 256: ...11 3 LR8501 Universal Unit Specifications 244...
Страница 262: ...12 3 Disposing of the HiLogger 250...
Страница 296: ...Index Index 4...
Страница 297: ......
Страница 298: ......
Страница 299: ......
Страница 300: ......