Assembly
89
Safety shield
➜
Move the flask lift upward
➜
Hang the safety shield to the edge of
the heating bath and position it so that
the (potentially present) safety hood
can be lifted open easily.
➜
Turn the eccentric screw 180° and
use to secure the safety shield on the
heating bath.
Safety shield
Heating bath
Eccentric screw
Inlet valve
(Optional accessories) Assembly parts:
▪
2x inlet hose
▪
1x inlet valve
During a running process, the sample flow
for the distillation is metered via the inlet
valve.
✓
The glassware set is under vacuum.
➜
Turn the inlet valve counterclockwise.
✓
The in-feed is opened, and the distillate
is drawn into the evaporating flask.
✓
The more you open the inlet valve, the
faster your sample will be supplied.
➜
Turn the inlet valve clockwise.
✓
The in-feed is reduced, or the inflow of
the sample is stopped.
Ventilation cap
Inlet valve
Mounting the inlet valve with the inlet hoses
All glassware sets (with the exception of glassware set G1) have a screw opening on the
lower left side of the condenser to accommodate the inlet hose with the inlet valve.