Assembly
82
Control box
The control box is mounted to the base
plate behind the heating bath.
➜
Attach the control box to the base
plate with the two included metric
screws.
Fixing bolt
The control box plug is on the rear of
the heating bath to the right, next to the
heating bath plug.
➜
Plug the cord of the control box into
the connection on the base plate.
Control box connection
Vacuum manifold
Assembly parts
▪
1x vacuum manifold
(holding plate with vacuum sensor,
ventilation valve and dirt filter)
▪
2x threaded screws
▪
1x torx wrench
The sheet of the vacuum manifold is
mounted onto the back side of the flask
lift.
➜
Use the torx wrench to screw the two
threaded screws into the back side of
the flask lift by a few turns.
➜
Hang the sheet of the vacuum
manifold onto the two threaded
screws.
➜
Tighten the two screws.
✓
The vacuum manifold is installed.
Threaded screws