Safety Instructions
8
▪
The device is designed for indoor use only under the following ambient conditions:
Ambient temperature
5 - 31 °C at 80% relative humidity
32 - 40 °C decreasing linearly to a
maximum 50% relative humidity
Installation altitude
0 - 2,000 m above sea level
Contamination level
2
Overvoltage category
II
Permissible supply
deviations
± 10 %
▪
If the device is operated in corrosive atmospheres, the service life of the device will
decrease depending on the concentration, duration and frequency of the exposure to
the corrosive atmosphere, e.g. concentrated hydrochloric acid (HCl)
Modifications to the device
▪
It is not permitted to make any unauthorized modifications or changes to
the device.
▪
Do not attach or install any parts that have not been approved by the manufacturer.
▪
Unauthorized modifications or changes will void the EC Declaration of Conformity for
the device and operation of the device will no longer be permitted.
▪
The manufacturer is not liable for any damage, dangers or injuries that result from
unauthorized modifications and changes or due to the non-observance of the instruc
-
tions in this manual.
Employee safety
➜
Ensure that the device is only operated by qualified specialist personnel and trained
employees.
➜
Comply with the following instructions to avoid any personal injuries and property
damage:
- Laboratory regulations
- Accident prevention regulations
- Hazardous Substance Act
- Other generally accepted occupational health and safety regulations
- Local regulations
Safety during use
➜
Wear the appropriate clothing when working on the device (protective clothing, safety
glasses and, if necessary, safety gloves).
➜
Do not use the device in potentially explosive areas. The device is not protected
against explosion. There is no explosion or ATEX protection available.
➜
Do not carry out work
with exposed flames in the vicinity of the device (risk of
explosion).
➜
Do not operate or assemble devices in the vicinity which are emission or radiation
sources (electromagnetic waves) for the frequency range (3*10
11
Hz to 3*10
15
Hz).
➜
Do not operate or assemble appliances in the vicinity of the device which constitute
emission or radiation sources for ionizing radiation or in the ultrasonic range.
➜
Do not operate the device where adiabatic compression or shock waves might occur
(shock wave combustion).
Содержание Hei-VAP Expert
Страница 198: ...Service 198...
Страница 199: ...199...