Troubleshooting
66
Error messages on display
Cause
Remedy
Information
Please close all openings on the
evaporator system and evacuate it.
When the pressure value has
become stable, enter the value
indicated on the external pressure
gauge and confirm it with OK!
The actual atmospheric
pressure value has
been entered
Enter/correct the actual
value of the
current minimum
vacuum pressure
Information
Vacuum sensor and/or
AUTOaccurate sensor not available!
Automatic mode (DAA) cannot be
started.
Vacuum sensor and/or
AUTOaccurate sensor
are not attached,
automatic mode (DAA)
cannot be started.
Attach vacuum sensor
and/or AUTOaccurate
sensor
Information
No ramp loaded
Ramp mode cannot be started!
No ramp profile loaded
▪
Load ramp profile
▪
Start ramp profile
mode
Resetting temperature limiter
➜
Switch the device off Remove the
cooled evaporating flask
➜
Disconnect the heating bath plug and
remove the heating bath.
➜
If necessary, empty the heating bath
and turn it over.
➜
Use a suitable object to briefly press
the red button on the bottom until it
clicks audibly into place.
✓
Reset the temperature limiter
Temperature limiter
If you experience a malfunction that cannot be resolved using the information listed
above, please inform your authorized Heidolph dealer immediately.