Troubleshooting
63
Error messages on display
Cause
Remedy
Error
A data bus error has occurred.
Switch the device off and on again
Error in data bus,
e.g. a bus node was
disconnected during
operation.
▪
Switch the device off
and on again!
Error
A temperature above the safety
temperature was measured in the
heating bath. The temperature
measurement unit is defective or
the medium filled in was too hot.
Switch the device off and on again!
Heating bath tempera-
ture is too high or the
temperature measure-
ment unit is defective.
▪
Let the heating bath
cool down
▪
Switch the device off
and on again
▪
Contact service
Error
An implausible temperature differ
-
ence was measured between the
heating bath sensors. The tempera-
ture measurement unit is defective.
Switch the device off and then on
again
Error in tempera-
ture measurement of
heating bath
Switch the device off
and then on again
Error
The temperature sensor for the
heating bath control is defective
(hardware monitoring unit).
Switch the device off and on again
Heating bath control is
defective, perhaps due
to short circuit
▪
Switch the device off
and on again
▪
Contact service
Error
The temperature sensor for the
heating bath control is broken or
the filled in bath fluid is too hot.
Switch the device off and on again
Heating bath control is
defective or tempera-
ture in the heating bath
>217°C
▪
Let the heating bath
cool down
▪
Switch the device off
and on again
▪
Contact service
Error
The temperature sensor for the
heating bath control has shorted or
the bath fluid was added too cold.
Switch the device off and on again
Heating bath control
is defective, perhaps
due to short circuit or
temperature <0 °C
▪
Heat up heating bath
▪
Switch the device off
and then on again
▪
Contact service
Error
The lift motor is defective or has
jammed!
Switch the device off and on again
Defective or jammed
lift motor
▪
Switch the device off
and on again
▪
Contact service
Error
The pressure exceeds the maximum
permissible excess pressure!
Pressure in the vacuum
system is too high
▪
Reduce pressure in
the vacuum system
Error
The set safety pressure was
exceeded.
Pressure in the vacuum
system is too high
▪
Reduce pressure in
the vacuum system
▪
If required, adjust
the set maximum
pressure