Troubleshooting
64
Error messages on display
Cause
Remedy
Error
The pressure sensor is not attached
or defective.
Pressure sensor not
attached or defective
or control box is not
connected correctly.
▪
Check and re-attach
vacuum sensor
▪
Check connection of
control box
Error
Defective communication with the
pressure sensor!
Hardware fault on the
vacuum plug-in board
▪
Contact service
▪
Replace vacuum
plug-in board
Error
The ventilation valve is not
connected or is defective.
Ventilation valve not
connected or defective
or control box is not
connected correctly.
▪
Check and re-connect
the ventilation valve
▪
Check connection of
control box
Error
The ventilation valve is defective.
Short circuit in the
ventilation valve
▪
Contact service
Error
The vacuum valve is not attached or
defective.
Vacuum valve not
attached or defective
or control box is not
connected correctly.
▪
Check and re-attach
vacuum valve
▪
Check connection of
control box
Error
The vacuum valve is defective.
Short circuit in the
vacuum valve
▪
Contact service
Error
The vacuum pump is not attached
or defective.
Vacuum pump not
attached or defective
or control box is not
connected correctly.
▪
Check and re-attach
vacuum pump
▪
Check connection of
control box
Error
Vacuum module:
Connection error between base
plate and vacuum plug-in board.
Please check the cable connection!
Bus error of the
component, e.g. due
to plugging/unplugging
during operation.
▪
Check cable
connection
▪
Switch the device off
and on again!
Error
Chiller module:
Connection error between base
plate and chiller plug-in board.
Please check the cable connection!
Bus error of the
component, e.g. due
to plugging/unplugging
during operation.
▪
Check cable
connection
▪
Switch the device off
and on again!
Error
A communication error between
base plate and control panel has
occurred!
Bus error of the
component, e.g. due
to plugging/unplugging
during operation.
▪
Check cable
connection
▪
Switch the device off
and on again!
Error
The rotary drive is overloaded or
defective!
Switch the device off and on again!
Drive motor defective
or overloaded
▪
Switch the device off
and on again!
▪
Contact service
Error
Alarm from the chiller:
Circuit board is too hot
Chiller overloaded
Switch chiller off and
let it cool down
Содержание Hei-VAP Expert
Страница 198: ...Service 198...
Страница 199: ...199...