*CV1?
*CV2?
*CV3?
*CV4?
*CV5?
*CV6?
Query the voltage of indicated channel.
Range: 0-150.0 Volts
Example: *123.45# indicates that the
control voltage is 123.45V for the
queried channel.
PMDC MODE
======================= =========================
*PMC STA?
Query PMDC operational status
*END#: PMDC search complete
*RUN#: PMDC is still running
*PMC:DGD OPT#
Scan available DGD values and optimize
polarization to maximize output DOP.
Scans through DGD index values 0-255.
*E00# if successful
See Table 2 for other error code
definitions.
*PMC:PMD OPT#
Scan available PMD values and optimize
polarization to maximize output DOP.
Scans through PMD index values 0-6560.
*E00# if successful
See Table 2 for other error code
definitions.
*PMC:DGD:CMP
idn
#
Optimize SOP to maximize output DOP at
specified DGD value (PMD ID#
idn
)
Range: 0-255
Ex: *PMC:DGD:CMP 16#
*E00# if successful
See Table 2 for other error code
definitions.
*PMC:PMD:CMP
idn
#
Optimize SOP to maximize output DOP at
specified PMD value (PMD ID#
idn
)
Range: 0 to 6560
Ex: *PMC:PMD:CMP 683#
*E00# if successful
See Table 2 for other error code
definitions.
*PMC CMP#
Optimize SOP to maximize output DOP at
current PMD setting (discrete or continuous)
*E00# if successful
See Table 2 for other error code
definitions.
MEMORY MODE
======================= =========================
*SAV:DSC:MB
X
loc,id
#
Discrete mode PMD sequence storage:
Store PMD ID#:
id
to memory location:
loc
,
in memory Bank X.
Range for loc: 0 to 99
Range for id: 0 to 6560
Range for X: A, B, C, D, or E
Example: *SAV:DSC:MBE 02,36#
Save PMD ID#:36 to memory location
02 in Bank E
*SAV:CON:MB
X
loc,dgd,so
#
Continuous mode PMD sequence storage:
Store DGD value
dgd
(in ps) and SOPMD
value
so
(in ps
2
) to memory location
loc
in
memory Bank X
Range for loc: 0 to 99
Range for X: A, B, C, D, or E
Example: *SAV:CON:MBA 2,5.90,205.0#
Save DGD 5.90ps, SOPMD 205.0ps² to
memory location 2 in Bank A.
*REC:MB
X
loc
#
Recall and generate PMD values stored at
memory location
loc
in memory bank X. If
value stored is a continuous setting (dgd, sopmd),
the PMDPro will recall the stored values, calculate
the closest available match, and generate the
calculated values.
Range for loc: 0 to 99
Range for X: A, B, C, D, or E
Ex: *REC:MBC 32#
Recall PMD stored at location 32 in
Memory Bank C and set PMDPro to
that value.
*REC:MB
X
loc
?
Query PMD values (continuous) or ID#
(discrete) stored at memory location
loc
in
memory bank X.
Will not set PMDPro to recalled PMD value.
Range for loc: 0 to 99
Range for X: A, B, C, D, or E
Continuous:
*CON,dgd,sopmd#
Example: *CON,25.5,180.0#
Discrete:
*DIS,id#
Example: *DIS,19#
Document #: GP-UM-PMD-1000-21
Page 99 of 122