Store/Recall:
This screen is used to store PMD sequences for scans or PMDC measurement results to
memory, or to recall stored data from memory.
PMD Memory Bank Store/Recall:
The PMD-1000 has 5 memory banks, with 100 locations each, for PMD sequence
storage. The user can thus store up to 5 sequences of PMD values which can be used for
the “Scan Stored PMD” function.
PMD sequence storage:
To store a PMD sequence, first select the desired memory bank (A-E).
Enter the values to be stored into the table at the right of the screen:
Sequence Number: Memory location index number
Discrete/Continuous (1/0): Indicates whether a discrete or a continuous mode
PMD value is to be stored at that location.
1 = Discrete (lookup table) value
0 = Continuous PMD value
PMD ID: If a discrete value is to be stored at the current location, type in the
index number of the desired DGD/SOPMD pair from the lookup table.
Index number range is 0-6560. For a discrete value, this fully specifies
the PMD. The DGD and SOPMD fields do not have to be filled in. For a
continuous PMD state, the PMD ID field does not have to be filled in.
DGD: For a continuous PMD state, type in the desired DGD, in ps.
SOPMD: For a continuous PMD state, type in the desired SOPMD, in ps
2
.
Document #: GP-UM-PMD-1000-21
Page 89 of 122