*PME:SET:PMD
idn
#
Discrete mode PMD control:
Set DGD/SOPMD to values corresponding to ID#
idn
Range: 0 to 6560.
*E00# if successful. See Table 2 for
other error code definitions. LCD
display will update.
PMDE MODE
===PMD Scans===
============================
*PME:SCN DGD
min,max
#
Discrete mode DGD scan:
Scan DGD values in lookup table from start ID
value
min
to end ID value
max
.
Range for
min
and
max
: 0 to 255; min <max.
*E00# if successful.
*E06# if values out of range.
See Table 2 for other error code
definitions. LCD display will update.
*PME:SCN DGD?
Query DGD scan limits
Range Boundaries: Min:0 and Max:255
DGD ID#s of start and end points
Example: *84,255#
*PME:SCN PMD#
Discrete mode PMD scan:
Scan through all PMD values in lookup table in
order of index number (i.e. from 0 to 6560)
*E00# if successful. See Table 2 for
other error code definitions. LCD
display will update.
*PME:CUR PMD?
Query Current PMD ID during a discrete mode
DGD or PMD scan.
*46#
*PME:SCN:SPD
time
#
Set the dwell time per point, in ms, for discrete
mode DGD or PMD scan.
Range: 1 to 60000
*E00# if successful. See Table 2 for
other error code definitions.
*PME:SCN SPD?
Query dwell time per point, in ms, for discrete mode
DGD or PMD scan.
*346#
*PME:SCN:CYC
cyc
#
Set the number of times (cycles) to run a scan
(Discrete scan or continuous/step trace)
Range: 1 to 9999
*E00# if successful. See Table 2 for
other error code definitions.
*PME:SCN CYC?
Query scan cycle number setting
*485#
*PME:SCN:CYC CUR?
Query Current Cycle count.
Number of completed cycles.
Example: *85#
*PME:SCN:CON ENA#
Begin Continuous Trace Scan using current settings. *E00# if successful. See Table 2 for
other error code definitions.
*PME:SCN:CON DIS#
Disable
(stop)
Continuous Trace Scan
*E00# if successful. See Table 2 for
other error code definitions.
*PME:SCN:STE ENA#
Begin
Step Trace Scan using current settings.
*E00# if successful. See Table 2 for
other error code definitions.
*PME:SCN:STE DIS#
Disable (stop)
Step Trace Scan
*E00# if successful. See Table 2 for
other error code definitions.
*PME:SCN:DWL
dwell
#
Set dwell time per point for step trace scan.
Range: 1 to 9999 ms
*E00# if successful. See Table 2 for
other error code definitions.
*PME:SCN DWL?
Query the Dwell time, in milliseconds
*234# i.e. 234 ms
*PME:SCN:DIR UPD#
For continuous or step trace scan:
Set the Scan Start Direction to UP
i.e. Scan from lower DGD limit to upper limit
*E00# if successful. See Table 2 for
other error code definitions.
*PME:SCN:DIR DND#
For continuous or step trace scan:
Set the Scan Start Direction to DOWN
i.e. Scan from upper DGD limit to lower limit
*E00# if successful. See Table 2 for
other error code definitions.
*PME:SCN DIR?
For continuous or step trace scan:
Query the scan direction.
Valid: *UPD#: UP-Direction
*DND#: DOWN-Direction
*END#: Scan done/ not scanning
*PME:SCN:MOD UNI#
For continuous or step trace scan:
Set the Scan MODE type to UNI-Directional
(one-way scan)
*E00# if successful. See Table 2 for
other error code definitions.
*PME:SCN:MOD BID#
For continuous or step trace scan:
Set the Scan MODE type to BI-Directional
(round-trip scan)
e.g. For direction “UP”, one cycle = round trip scan
from lower DGD limit to upper limit and back.
*E00# if successful. See Table 2 for
other error code definitions.