*SAV:PMC
loc
#
Store the most recent PMDC Search result to
PMDC memory location
loc
.
Range for loc: 0 to 99
Ex: *SAV:PMC 14#
*REC:PMC
loc
#
Recall the stored PMDC Search result from
PMDC memory location
loc
and set PMDPro
to the recalled setting.
Range for loc: 0 to 99
*E00# if successful.
*E12# if location is empty or contains
invalid data.
See Table 2 for other error code
definitions.
SETUP MODE
========================= ==========================
*CHA?
Query ITU grid channel number setting.
See Appendix A for wavelength/frequency
correspondence.
L Band Range: L001 to L130
C Band Range: C001 to C087
Example : *L112#
or *C012#
*CHL nnn# Set L-BAND
*CHC nnn# Set C-BAND
Set wavelength/frequency channel setting
using ITU grid index#- see Appendix A
L Band Range: L001 to L130
C Band Range: C001 to C087
*E00# if successful. See Table 2 for
other error code definitions.
Available range depends on PMDPro
configuration ordered.
*ADR
adr
#
Set GPIB address
Range: 1 to 30
*E00# if successful. See Table 2 for
other error code definitions.
Example: *ADR 5#
*ADR?
Query GPIB address
Return : *30#
Set Ethernet Mode and
Query IP Address
Note: This set of commands must be sent
from RS-232 only
~7DHCP#
Set instrument to dynamic IP mode
OK
~6STIP#
Set instrument to static IP mode
OK
~8IP?# Request
IP
address
(Note: Returns either dynamic or static IP
address, depending on which mode is active.)
(IP Address will be returned)
Table 2 Error Code Definitions
E00
No error (Correctly formatted command received)
E01 Undefined
Command
E06
Parameter outside the allowed range
E07
String of characters too long(>buffer limit)
E08
Input power too high
E09
Input power too low
E10
Option Not Supported when in Continuous Mode
E11
ID# out of range
E12
Memory data location is empty or stored data is
invalid
E13
Option Not Supported when in Discrete Mode
E50
Invalid Stokes parameter values
Sum of squares of Stokes values
≠
1
Document #: GP-UM-PMD-1000-21
Page 100 of 122