4.
GP_SET_ETHERNET program:
To run this program:
a)
Connect the PMDPro to the control computer with a straight wired RS-232
cable as described in section 4.3.3. This program is RS-232 only, and cannot
be run from any other interface.
b)
Set the control computer’s baud rate to 28800.
c)
Open the program GP_SET_ETHERNET.Select the correct serial (RS-232)
port from the port configuration box at the top of the screen. The baud rate
should be set to 28800. Once these two parameters are set, click the white
arrow to run the program.
The program has 3 tabs.
The first tab, “GET IP CONFIG” is used to query the current IP configuration of the
instrument. When the “Get IP Config” button at the left of the screen is clicked, the
program will read back the instrument’s current network parameters, as well as
whether the instrument is currently set for static or dynamic IP addressing.
IP Config query screen- Static IP mode result
Document #: GP-UM-PMD-1000-21
Page 52 of 122