18
Table 1:
A general guide for the chemical resistance of the wetted parts in the
development unit.
Wetted parts
1
Material ofGenerally
Generally
construction
resistant to
attacked by
Distributor and
PVDF
2
strong acids and
ketones, esters, and hot
distributing plate
PVDF
2
bases in moderate acids
concentration and
alcohols and
hydrocarbons
Gasket (10-port
fluoro rubber
moderate acids,
hot strong acids, esters,
valve)
strong bases,
ketones, and bleach
many solvents,
alcohols,
aldehydes
Tubin (10-port
Teflon
most chemicals
extreme conditions
to chamber)
Tubing (bottles
PVC
3
strong acids and
hot acids, ketones, and
to 10-port valve)
bases in moderate hydrocarbons
concentration,
alcohols,
aldehydes, and
bleach
Chamber, gel
stainless
most chemicals
long exposure to salt
holder, and temp.
steel
solutions
sensor
Chamber lid gasket
EPDM
4
strong acids and
hot acids and aromatic
bases, alcohols,
hydrocarbons
aldehydes, and
ketones
Chamber lid
PP
5
strong acids in
hot strong acids, aromatic
moderate con-
hydrocarbons, and bleach
centration in high
concentration,
alcohols, alde-
hydes, and
ketones
1
These parts are illustrated on pages 16, 17 and 79.
4
Ethylene propylene copolymer and terpolymer
2
Polyvinylidine fluoride
5
Polypropylene (or polypropene)
3
Polyvinyl chloride
At present, PhastGel media are available for four types of electrco-
phoretic techniques; native polyacrylamide gel electrophoresis
(PAGE) in gradient or homogeneous gels, SDS-PAGE in gradient or
homogeneous gels, and isoelectric focusing (IEF). These gels can be
combined for twodimensional techniques.
PhastGel media are made of polyacrylamide bonded to a transparent
polyester backing. The gel surface is covered with a plastic film which
prevents drying and contamination. This film must he peeled off
directly before use, that is, after the gel has been positioned onto the
separation bed, and excess water has been removed.
PhastGel media are individually packaged in airtight envelopes. Once
removed from their package, gels should be used immediately.
PhastGel chemicals include buffer strips for electrophoresis and a
Coomassie-type stain, PhastGel Blue R. PhastGel media and chemi-
3.3 PhastGel media
and chemicals
3. Description of the system
Содержание PhastSystem
Страница 1: ...Phast System user manual automated electrophoresis um 80 1320 15 Edition AI ...
Страница 2: ......
Страница 8: ...8 ...
Страница 30: ...30 ...
Страница 34: ...34 ...
Страница 64: ...64 ...
Страница 96: ...12 ...
Страница 104: ......
Страница 105: ......
Страница 106: ...PRINTED IN SWEDEN BY TK I UPPSALA AB 2003 ...