33
Mains connection
Plug the mains power cords (120 V or 220 V) into the input marked
MAINS on the rear panel of the units. Plug the cords into the wall
outlet (grounded to earth).
Important!
Always disconnect these cords when servicing the
instruments.
The system is turned on by pressing in the on/off button on the rear
panel of the separation and control unit. The development unit is
automatically activated when a development method is started.
Diagnostics
Turn on the system and check that the diagnostics are successfully
completed. PhastSystem does a self-diagnostic test every time it is
turned on. If an error is detected during the test, a message will
appear on the display and an alarm will sound.
Temperature sensors
Both units have a temperature sensor; one is under the separation bed
and the other is on the underside of the lid in the development
chamber (enclosed in stainless steel). These are calibrated before
shipment, but you may want to check them before using
PhastSystem. The sensors can be checked and calibrated individually.
See the chapter on Maintenance for instructions.
Before using the development unit we recommend that you run a
cleaning method to remove dust accumulated during storage and
shipment. A cleaning method requires only distilled water and the
level sensor shield (instead of gels). The level sensor shield is in the gel
holder in the development chamber when you receive PhastSystem.
See chapter 5, the section on cleaning method, for instructions.
The level sensor shield must remain in the chamber when running
methods without gels; otherwise, the chamber will not fill.
Warning!
The level sensor (on the underside of the development
chamber lid) is enclosed in glass and is quite fragile. Use
extreme care when cleaning this sensor.
4.3 Turning the
system on
4.4 Before use
4. Installation
Содержание PhastSystem
Страница 1: ...Phast System user manual automated electrophoresis um 80 1320 15 Edition AI ...
Страница 2: ......
Страница 8: ...8 ...
Страница 30: ...30 ...
Страница 34: ...34 ...
Страница 64: ...64 ...
Страница 96: ...12 ...
Страница 104: ......
Страница 105: ......
Страница 106: ...PRINTED IN SWEDEN BY TK I UPPSALA AB 2003 ...