6
211>POWER FAILURE – METHOD(S) SET TO PAUSE
If mains power fails longer than 5–10 seconds,
running methods will be paused. When power
returns, an alarm will sound to inform you
about the power failure. You must press ”SEP
pause/continue” or ”DEV pause/continue”
for the method(s) to continue from where they
left off.
When mains power fails less than 5–10 sec-
onds, runnings methods will automatically
continue from where they were stopped once
power is returned.
212>TEMP. SENSOR NOT CALIBRATED
Calibrate the temperature sensors according
to the instructions given on page 75 of the
manual.
System error messages
The following messages will appear on the
display if an error is detected during diagnostics
when you turn the system on:
301>BATTERY NOT CONNECTED
302>BATTERY VOLTAGE TOO LOW
303>SEPARATION TEMP. SENSOR NOT OK
304>DEVELOPMENT TEMP. SENSOR NOT OK
305>PROM CHECKSUM NOT OK
306>PROCESSOR INTERNAL RAM NOT OK
307>EXTERNAL RAM NOT OK
308>PROCESSOR TIMERS NOT OK
309>AD-CONVERTER NOT OK
First try turning the system on and off several
times. Fluctuations in the main power may cause
a false error to appear on the display. If the error
message still appears, note down the message and
call for service.