8
Trouble Shooting Guide
Probably Cause
Solution bottles were incorrectly
connected to the ports according to
the programmed method, or, the
wrong method was started.
Finger proteins from handling the gel
without using forceps or gloves.
This usually occurs when gels are
dried too fast e.g. the air stream from
the hair-dryer is too hot or too high.
Solution(s)
Always label bottles with their corre-
sponding port number. Label tubes
with their corresponding port num-
bers using the yellow tubing mark-
ers. Check the program against the
bottle arrangement.
Use the plastic-tab from the gel back-
ing to handle the gels. Use forceps.
Use the lowest heat setting on the
hair-dryer.
Symptoms
No bands but blue
background
Blue blotches near
the gel edges
Small cracks
and rough
uneven
surface after
drying