3
Programming error messages
If an alarm sounds while you are programming or
editing a method, or if an alarm sounds when you
start a method, press ”help/return” to find out
what the problem is. One of the following
messages will be displayed:
101>ILLOGICAL SAMPLE APPLICATOR VALUES
When starting a separation run with a method
which has illogical sample applicator values,
for example, if the sample applicator is pro-
grammed to go down after 22 Vh in step 1 and
step 1 only has 20 Vh
102>METHOD = 0 DOES NOT EXIST
When you try to call up method 0; methods
are numbered from 1- 9.
103>ILLOGICAL METHOD STEP PARAMETERS
When trying to start a separation method with
illogical parameters other than applicator
movement, for example, 0 V.
104>DEV PAUSE PROHIBITED IF DEV OFF
When trying to pause a development run not
started.
105>SEP PAUSE PROHIBITED IS SEP OFF
When trying to pause a separation run not
started.
106>INSERT ALLOWED ONLY WITHIN METHOD
When trying to insert before entering a
method in the program mode.
107>TO EDIT A RUNNING METHOD DO <pause>
When trying to edit a running method without
pressing ”SEP pause/continue” or ”DEV
pause/continue” first.
108>source method does not exist
When trying to copy a free (not programmed)
method.
109>METHOD STEP = 0 DOES NOT EXIST
When trying to call up step 0; steps are num-
bered 1 to 9 in separation methods and 1 to 20
in development methods.
110>INSERT ALLOWED ONLY AT METHOD STEPS
When trying to insert at an applicator or
alarm instruction in a separation method, or
at a Ct or alarm instruction in a development
method.
111>NO FREE STEPS AFTER PRESENT
When trying to insert a step in a method when
all subsequent steps are programmed. You
may insert if you first delete one of the subse-
quent steps.
112>DESTINATION METHOD OCCUPIED
When trying to copy a method to a pro-
grammed method.
113>COPY FROM METHOD TO STEP PROHIBITED
When trying to copy a whole method into one
step.
114>SOURCE METHOD STEP DOES NOT EXIST
When trying to copy a step that is not
programmed.
115>DESTINATION METHOD STEP OCCUPIED
When trying to copy to a step that is already
programmed.
116>DELETE OF RUNNING METHOD PROHIBITED
When trying to delete a running method. You
must stop the method first before you delete it.
117>CANNOT DELETE A RUN/RUNNING STEP
When trying to delete a step that has run or a
step that was running when the method was
paused. Steps can be deleted in a running
method only if they follow the paused running
step.
118>TO INSERT A NEW STEP DO
<pause>
When trying to insert a step in a running sepa-
ration or development method without first
pressing ”SEP pause/continue” or ”DEV
pause/continue”.
Help Message Reference
Содержание PhastSystem
Страница 1: ...Phast System user manual automated electrophoresis um 80 1320 15 Edition AI ...
Страница 2: ......
Страница 8: ...8 ...
Страница 30: ...30 ...
Страница 34: ...34 ...
Страница 64: ...64 ...
Страница 96: ...12 ...
Страница 104: ......
Страница 105: ......
Страница 106: ...PRINTED IN SWEDEN BY TK I UPPSALA AB 2003 ...